Specifications

Chapter 6 152
Test Parameters
Parameter descriptions
29 RX Usable Input Level Range RF Level
This parameter sets the HP/Agilent 8922 RF generator’s amplitude
level to be used during input level range testing. The range is
19 to
127 dBm.
Default value The GSM900 default RF level for input level range testing is
15 dBm
The DCS1800 default RF level for input level range testing is
23 dBm
The PCS1900 default RF level for input level range testing is
23 dBm
NOTE
The default value of parameter 29, RX usable input level range
RF level for GSM is -15 dBm which adheres to the ETSI
specifications. However, this default value does not run on the HP/
Agilent 8922E or G. If you have an HP/Agilent 8922E or G, you must
either create a procedure with the value of parameter 29 set to -19
dBm, or upgrade your HP/Agilent 8922 to a model S, M, R or P. For
more details on how to create your own test procedures, see “Making
your own test procedures” on page 24.
Example If you want the RF level set to
20 dBm when testing the receiver’s
usable-input level range, you would enter
20 in the value field.
NOTE
The HP/Agilent 8922E and HP/Agilent 8922G are specified to a
maximum RF Generator amplitude limit of
19 dBm.
The HP/Agilent 8922F and HP/Agilent 8922H are specified to a
maximum RF Generator amplitude limit of 13 dBm.
The Agilent 8922S and Agilent 8922M are specified to a maximum RF
Generator amplitude limit of 14 dBm.
30 RX Bits to Test Usable Input Level II
This parameter allows you to enter the number of type II bits to test
during receiver usable input level testing. The range is 0 to 99999999.
Default value The default type II bits to test is 10000.
Example If you want to test usable input level using 40,000 type II bits, enter
40000 in the value field.