Test Specifications
37 TX Worst Case Phase Error (peak)
37 TX Worst Case Phase Error (peak)
Description This specification sets the upper limit for the mobile station
transmitter’s worst case phase error, expressed in degrees (peak).
NOTE
”Worst case” phase and frequency error measurements only apply when
parameter TX phase/freq measurement averages is set to
something other than its default value of 1.
If the limit is exceeded, an “F” (for fail) is displayed next to the TX wc
phase error peak measurement.
Default value The default value for the upper limit is 20
° peak.
Example If you want your test procedure to report a failure if the transmitter
worst case phase error exceeds 25 degrees you would enter 0 in the
lower-limit column and 25 in the upper-limit column.