Specifications

Chapter 5 115
Test Specifications
6 TX Burst Timing Measurement
6 TX Burst Timing Measurement
Description This specification sets the mobile station’s burst timing error limits.
If a selected limit is exceeded, an “F” (for fail) will be displayed next to
the TX timing error measurement result.
Default values The default value for the lower limit is
1 T (bit period). The default
setting for the upper limit is 1 T (bit period).
Example If you want your test procedure to report a failure if the burst timing
error of the mobile station exceeds
±2 T, you would enter 2 in the
lower-limit column and 2 in the upper-limit column. Limits are entered
in units of “T” (bit period).