Specifications
Chapter 5 108
Test Specifications
Default specifications
Default specifications
The default specifications come from the GSM 11.10 recommendations
for Class 5 mobile stations, GSM 11.10-DCS recommendations and the
document PN 3389 Personal Communications Services Air Interface
Specification for Class 2 mobile stations.
When a test procedure is run, an “F” (for fail) is displayed next to each
measurement result that exceeds its limits from the specifications file.
Any combination of upper and lower limits can be selected in the Check
column. See “TESTS (Edit Specifications)” on page 170.
The test procedure stops when a specification value is exceeded if the
Stop on failure field is set to Yes. See “Test execution conditions”
on page 23.
NOTE
A Test Procedure must be selected before the Specifications list
appears.
NOTE
Specifications 8 – 16 and 23 only apply to the HP/Agilent 8922 with
Spectrum Analyzer option.
NOTE
Certain specifications are hardcoded into the Agilent 83212D software
and are not user editable. This makes it easier to define tests and
removes scope for error due to misconfiguration.