Specifications
Chapter 4 94
Tests
TEST_10 RX Usable Input Level Range
TEST_10 RX Usable Input Level Range
Description This test verifies the dynamic range of the mobile station’s receiver by
performing bit error testing at a high signal level. The default level
transmitted to the mobile station is
−15 dBm, set by the parameter 29,
RX usable input level range RF level.
NOTE
The default value of parameter 29, RX usable input level range
RF level for GSM is -15 dBm which adheres to the ETSI
specifications. However, this default does not run on the HP/Agilent
8922 model E or G. If you have an HP/Agilent 8922 E or G, you must
either create a procedure with the value of parameter 29 set to -19
dBm, or upgrade your HP/Agilent 8922 to a model S, M, R or P. For
more details on how to create your own test procedures, see “Making
your own test procedures” on page 24.
Test results display the bit error count or ratio and pass/fail result for
each channel selected. Receiver usable input level range is measured in
count, percent, or ppm error (depending upon how the RX BER/FER
results parameter is set).
Specification
used
Parameters
Propagation Type Max Events Max Samples
Static (Class II) 200 (0.0122%) 1,640,000
Spec
no.
Description Units GSM900
Default
Lower
Limit
GSM900
Default
Upper
Limit
DCS1800
Default
Lower
Limit
DCS1800
Default
Upper
Limit
PCS1900
Default
Lower
Limit
PCS1900
Default
Upper
Limit
5 RX usable
input level
type II BER
ct/%/
ppm
0.1 0.1 0.1
Parm
no.
Description Units GSM900
Default
Value
DCS1800
Default
Value
PCS1900
Default
Value
3 CP base station color code code no. 5.0 5.0 5.0
4 CP public land mobile network color
code
code no. 1.0 1.0 1.0
5 CP serving cell (BCH) ARFCN chan. number 25.0 600.0 600.0