Specifications

Chapter 4 90
Tests
TEST_08 TX ORFS Due to Ramping
Specifications The following hardcoded values are used for specifications 13 to 16:
NOTE
Default specifications have been adjusted for a 3-pole filter instead of
the 5-pole filter indicated in the GSM recommendations. These values
are for Class 5 GSM and Class 2 DCS/PCS mobile stations. Refer to the
technical reference sheet Output RF Spectrum Measurements Using a
3-Pole Synchronously Tuned Measurement Filter (Agilent part number
1000-1223) available from the nearest Regional Sales and Service
Office.
Frequency offset
100 kHz 200 kHz 250 kHz 400 kHz 600 - 1600
kHz
GSM +0.5 dBc or
-36 dBm
-27.5 dBc or
-34.5 dBm
-32 dBc or
-35 dBm
-51 dBc or
-27 dBm
-60 dBc or
-51 dBm
DCS/PCS
-60 dBc or
-56 dBm