User`s guide

2-53
GPIB Tutorial and Examples
Transient Settling Times
e) Arming DSP Analyzer, Output RF Spectrum or Pulse On/Off measurements in
Single mode from IBASIC - after sending the Arm command wait approximately 5
seconds.
OUTPUT 714,”TRIG:ASTate ’ARM’”
WAIT 5
f) IMEI Request: 10 seconds
OUTPUT 714,”MSIN:MS:IMEI:REQ”
- WAIT 10
g) TMSI Reallocation: Query TMSI value (should change within 10 seconds)
i. Read TMSI string OUTPUT 714,”MSIN:PAG:TMSI?”
ii. Send TMSI Reallocation command OUTPUT
714,”MSIN:PAG:TMSI:REAL”
iii. Wait until OUTPUT 714,”MSIN:PAG:TMSI?” returns a new string
1 10 seconds maximum
We expect users, operating remotely, to make measurements in single mode, mainly for
speed reasons. However, if you operate in continuous (CONT) measurement mode and
you change a parameter that affects the measurement result, then (at a maximum) the third
measurement result queried will be an outcome of the new setup and not the previous
setup.