User`s guide

D-3
Appendix D - [:MM] - Measurement Fields
Commands
Commands
STATe STATe?
STATeON | OFF | 1 | 0
Selects/queries the STATe of the measurement to be ON or OFF.
Note: ON = 1 and OFF = 0
UNITs UNITs?
UNITs <GPIB measurement unit>
Sets/queries the GPIB fundamental UNITs that measurement queries will be returned in.
DUNits DUNits?
DUNits <measurement unit>
Sets/queries the Displayed UNits on the front panel for the given measurement.
AUNits AUNits?
AUNits <GPIB attribute unit>
Sets/queries the GPIB fundamental UNITs that measurement Attribute queries (e.g., low
limit, high limit, etc.) are returned in.
HLIMit HLIMit
High LIMit measurement information.
HLIMit:
DUNits
HLIMit:DUNits?
HLIMit:DUNits <measurement unit>
Sets/Queries the measurement High LIMit Displayed UNits.
HLIMit:
EXCeeded
HLIMit:EXCeeded?
Queries whether the High LIMit for the measurement was EXCeeded.
HLIMit:
RESet
HLIMit:RESet
RESet the High LIMit exceeded state so that new limit data can be acquired.
HLIMit:
STATe
HLIMit:STATe?
HLIMit:STATe ON | OFF | 1 | 0
Sets/queries the High LIMit STATe. Note: ON = 1 and OFF = 0.
HLIMit
[:VALue]
[:VALue]?
[:VALue] <numeric value>
Sets/queries the measurement High LIMit VALue.
LLIMit LLIMit
Low LIMit measurement information.