User`s guide

40-8
Trigger Subsystem
MODE[:DSP]
MODE[:DSP]
Description Selects/queries the DSP TRIGger MODE as SINGLE or CONTinuous. This is used for
Phase, Amplitude, Output RF Spectrum, Pulse On/Off Ratio and Data Bits measurements.
Syntax TRIGger:MODE[:DSP]?
TRIGger:MODE[:DSP] <string>
Options ’SINGLE’ | ’CONT’
NOTE In CONTinues mode, the user does not manually arm the instrument, but must provide a
trigger in order for the measurement to complete.
NOTE This command is valid in both local and remote modes.
MODE:RETRigger
Description Selects/queries the RETRigger MODE for the currently active measurement. Default
setting is REPetitive.
Syntax TRIGger:MODE:RETRigger?
TRIGger:MODE:RETRigger <string>
Options ’SINGLE’ | ’REPETITIVE’
CAUTION: The remote-only command will override local triggering commands for continuous
(repetitive) and single settings for AF Analyzer, CW Meas, OSCilloscope, and SANalyzer.