User`s guide
39-6
Tests Subsystem
EXECution:RESults
EXECution:RESults
Description Sets/Queries the output results sent to the output device (CRT/Printer).
Syntax TESTs:EXECution:RESults?
TEST:EXEC:RES?
TESTs:EXECution:RESults <string>
TEST:EXEC:RES <string>
Options ‘ALL’ | ‘FAILURES’
Where;
ALL All test results are shown on the output device (CRT and/or printer).
Printouts include a “banner” listing the test conditions, measured
values, lower and upper limits, and whether the test passed or failed.
The Comment field is shown at the top along with any identifying
information from the Output Heading field. Date, and time is
also output.
FAILURE Test results are shown only when a UUT failure or software error
occurs. Printouts include a “banner” listing the test conditions,
measured values, and lower and upper limits of the failed test. The
Comment field and any identifying information from the Output
Heading field is also output.