User`s guide
37-3
Status Subsystem
Status Subsystem - Status Byte
Status Subsystem - Status Byte
The Status subsystem is used for setting and querying the various conditions of the
instrument through the conditions set within the status byte. The following is a description
of the states found with each of the parts within the status byte.
Status Byte Bit Definitions
7 - Operation Status Register
6 - RQS
5 - Standard Event Status Register
4 - MAV
3 - Questionable Data/Signal Status Register
1 - Hardware 2 Status Register
0 - Hardware 1 Status Register
Hardware 1 Status Register
Condition register bit
deļ¬nitions
7 - Communication Status Register Summary
6 - Power up tests failed
5 - Pulse On Trace RF Overload
(PULSe)
4 - Measurement Trigger too early
(DSPanalyzer, ORFSpectrum, PULSe)
3 - Measurement Trigger too late
(DSPanalyzer, ORFSpectrum, PULSe)
2 - Measurement Sync Error
(DSPanalyzer, ORFSpectrum, PULSe)
0 - Measurement armed
(DSPanalyzer, ORFSpectrum, PULSe)