User`s guide

32-5
RF Analyzer Subsystem
AMPLitude2
AMPLitude2
Description Sets/queries the amplitude (input level to assume) of the AUX RF IN port. Used when
RFAN:INP is ’AUX RF IN’.
GPIB and display units are dBm, Volts (V) and Watts (W);
Default GPIB and display unit is dBm.
Syntax RFANalyzer:AMPLitude2?
RFANalyzer:AMPLitude2 <real> | [:FNUM]
Options Refer to Appendix B.
[:AMPLitude]:ACCuracy
Description Selects/queries the RF ANalyzer AMPLitude ACCuracy.
Syntax RFANalyzer[:AMPLitude]:ACCuracy?
RFANalyzer[:AMPLitude]:ACCuracy <string>
Options ‘+-3dB’ | ‘+-1dB’
[:AMPLitude]:CONTrol
Description Selects/queries the RFAnalyzer AMPLitude CONTrolling mechanism.
Syntax RFANalyzer[:AMPLitude]:CONTrol?
RFANalyzer[:AMPLitude]:CONTrol <string>
Options ’MS TX LEV’ | ’MANUAL
Where;
MS TX LEV means that the AMPLitude (RFAN:AMPL1 or RFAN:AMPL2) is set
automatically based on the setting of CELL:MS:TLEVel.
MANUAL means that the user can manually set the AMPLitude (RFAN:AMPL1 or
RFAN:AMPL2)