User`s guide
32-4
RF Analyzer Subsystem
AGC:CALibrate
AGC:CALibrate
Description Does an open loop AGC CALibration if in FRAN:AGC:MODE ‘CLOSED’.
Syntax RFANalyzer:AGC:CALibrate
Options Not Applicable.
AGC:DVALue
Description Sets/queries the open/auto AGC DAC VAlue
Syntax RFANalyzer:AGC:DVALue?
RFANalyzer:AGC:DVALue <integer> | [:INUM]
Options Refer to Appendix A.
AGC:MODE
Description Selects/queries the AGC MODE (NORMAL is closed loop).
Syntax RFANalyzer:AGC:MODE?
RFANalyzer:AGC:MODE <string>
Options ’CLOSED’ | ’OPEN’ | ‘AUTO’
AMPLitude1
Description Sets/queries the amplitude (input level to assume) of the RF IN/OUT port. Used when
RFAN:INP is ’RF IN/OUT’.
GPIB and display units are dBm, Volts (V) and Watts (W);
Default GPIB and display unit is dBm.
Syntax RFANalyzer:AMPLitude1?
RFANalyzer:AMPLitude1 <real> | [:FNUM]
Options Refer to Appendix B.