User`s guide

31-5
Pulse On/Off Ratio Commands (Measure Subsystem)
MARKer:TIME:RISE
MARKer:TIME:RISE
Description Queries the RISE trace TIME which is the marker’s position relative to bit zero in the
measured burst. This value is a function of the marker position set or queried by
PULS:MARK:POS:RISE.
GPIB units are seconds (S),bit periods (T);
default unit is seconds (S).
Display units are US (micro-second), T (bit periods);
default unit is US (micro-second).
Syntax MEASure:PULSe:MARKer:TIME:RISE?
MEASure:PULSe:MARKer:TIME:RISE[:MM] | [:AVG]
Options Refer to Appendices D and F.
OORatio:FALL
Description Queries the PULSe On/Off Ratio FALL trace MEASurement result.
Default units: dB relative to the average power over the useful bits in the measured burst.
GPIB unit is dB.
Display unit is dB.
Syntax MEASure:PULSe:OORatio:FALL?
MEASure:PULSe:OORatio:FALL[:MM] | [:AVG]
Options Refer to Appendices D and F.
OORatio:RISE
Description Queries the PULSe On/Off Ratio RISE trace MEASurement result.
Default units: dB relative to the average power over the useful bits in the measured burst.
GPIB unit is dB.
Display unit is dB.
Syntax MEASure:PULSe:OORatio:RISE?
MEASure:PULSe:OORatio:RISE[:MM] | [:AVG]
Options Refer to Appendices D and F.