User`s guide

27-3
Oscilloscope Commands (Measure Subsystem)
MARKer:LEVel:AM
MARKer:LEVel:AM
Description Queries the MARKer LEVel which is the signal level of the current marker position for AF
Analyzer input selections (AFAN:INP) that have AM units of Percent. This value is a
function of the marker position set or queried by OSC:MARK:POS.
GPIB unit is Percent (PCT);
Display unit is Percent (PCT).
Syntax MEASure:OSCilloscope:MARKer:LEVel:AM?]
MEASure:OSCilloscope:MARKer:LEVel:AM[:MM] | [:AVG]
Options Refer to Appendices D and F.
MARKer:LEVel:FM
Description Queries the MARKer LEVel which is the signal level of the current marker position for AF
Analyzer input selections (AFAN:INP) that have FM units of Hertz. This value is a function
of the marker position set or queried by OSC:MARK:POS.
GPIB units are HZ, kHZ;
Display units are kHZ.
Syntax MEASure:OSCilloscope:MARKer:LEVel:FM?]
MEASure:OSCilloscope:MARKer:LEVel:FM[:MM] | [:AVG]
Options Refer to Appendices D and F.
MARKer:LEVel:VOLTs
Description Queries the MARKer LEVel which is the signal level of the current marker position for AF
Analyzer input selections (AFAN:INP) that have units of VOLTs. This value is a function
of the marker position set or queried by OSC:MARK:POS.
GPIB unit is Volts (V);
Display units are V, mV
default unit is V.
Syntax MEASure:OSCilloscope:MARKer:LEVel:VOLTs?]
MEASure:OSCilloscope:MARKer:LEVel:VOLTs[:MM] | [:AVG]
Options Refer to Appendices D and F.