User`s guide
21-8
IEEE 488.2 Common Commands
*RCL (Recall)
*RCL (Recall)
Description The *RCL command restores the state of the instrument from the
specified internal save/recall register. An instrument setup must have
been stored previously in the specified register. Registers 0 through 99
are general purpose and can be used with the *SAV command.
Command Syntax *RCL <rcl_register>
Where <rcl_register> = 0 through 99 though the total number of
registers used may be limited by the amount of memory available.
Example
OUTPUT 714;”*RCL 75”
An instrument state stored using [REGister:]SAVE may be recalled
using *RCL or [REGister:]RECall. If the [REGister:]SAVE uses an
alphanumeric string as the register name, the *RCL command will not
work. *RCL only works with registers named using an integer from 0
through 99.
IMPORTANT The following fields do not participate in Save/Recall, and will be set according to the
recalled state of the operating mode (Active Cell | Test Mode | CW Generator).
• Cell Config - Settable| | Activated
• Dig Demod Arm State - Arm | Disarm
• DSP Meas - Trig Source
• DSP Meas - Trig Delay
• Demod Arm State - Arm | Disarm
• Meas Arm State - Arm | Disarm
• Meas Sync - Single | Cont
• Bit Error Test - Run | Stop
• Bit Error Test MS Loopback Loop Delay mode - Manual | Auto
• Hop Control RF Generator - Arm | Disarm
• Hop Control RF Analyzer - Arm | Disarm
• Hop Control RF Generator - Non-Hop | Hop
• Hop Control RF Analyzer - Non-Hop | Hop
• None of the CONFigure commands except :RADio,
:ROSCillator:OFFSet, ROSCillator:TUNing, and :PRINt:TITLe
participate in Save/Recall, and will instead remain at their last
setting.