21-6
IEEE 488.2 Common Commands
*OPC (Operation Complete)
*OPC (Operation Complete)
Description The *OPC (operation complete) command will cause the instrument
to set the operation complete bit in the Standard Event Status Register
only when all pending operations are complete. The *OPC? query
places an ASCII “1” in the output queue when all pending device
operations are complete. There is a one second minimum delay
between the query and the response. A pending operation in the
Agilent 8922M or Agilent 8922S is any measurement which is armed
but not complete. When in remote operation with repetitive triggering
all measurements, apart from BER, are self-arming. When in remote
operation with single triggering all measurements, apart from BER,
are armed by sending the TRIGger[:IMMediate] command or
*TRG. The BER measurement is armed by sending the
TRIGger:BETest:RUN command.
Command Syntax *OPC
Example
OUTPUT 714;”*OPC”
Query Syntax *OPC?
Returned Format
1<NL>
Example
OUTPUT 714;”*OPC?”
ENTER 714;Op
PRINT Op