Hop Control Subsystem
RFANalyzer or RFGenerator:INSert
RFANalyzer or RFGenerator:INSert
Description This INSerts an entry of 0 MHz into the RF ANalyzer or RF Generator hop table. All other
entries move down one address. The last entry in the RF ANalyzer or RF Generator hop
table is lost.
Syntax HOPControl:RFANalyzer:INSert
HOPControl:RFGenerator:INSert
Options Not applicable.
RFANalyzer or RFGenerator:MODE
Description Selects/queries the RF ANalyzer or RF Generator hop MODE.
Syntax HOPControl:RFANalyzer:MODE?
HOPControl:RFGenerator:MODE?
HOPControl:RFANalyzer:MODE <string>
HOPControl:RFGenerator:MODE <string>
Options ‘NON-HOP’ | ‘HOP’
RFANalyzer or RFGenerator:SETTling
Description Selects/queries the RF ANalyzer or RF Generator hop SETTling.
Syntax HOPControl:RFANalyzer:SETTling?
HOPControl:RFGenerator:SETTling?
HOPControl:RFANalyzer:SETTling <string>
HOPControl:RFGenerator:SETTling <string>
Options ‘NORMAL’ | ‘LARGEHOPS’
Where;
NORMAL should be used for small hops.