User`s guide
20-6
Hop Control Subsystem
RFANalyzer or RFGenerator:CTENd
RFANalyzer or RFGenerator:CTENd
Description This Clear To ENd command replaces the RF ANalyzer hop frequency entry at
HOPC:ADDRess, and all hop frequency entries after it, with 0 MHz.
For example;, the hop frequency entry at HOPC:ADDRess, HOPC:ADDRess + 1, ... up to
entry 2047 are replaced with 0 MHz
Syntax HOPControl:RFANalyzer:CTENd
HOPControl:RFGenerator:CTENd
Options Not applicable.
RFANalyzer or RFGenerator:DELete
Description This DELetes the RF ANalyzer hop frequency entry at HOPC:ADDRess. All other RF
ANalyzer hop frequency entries move down by one address. Entry 2047 in the hop table is
replaced with 0 MHz.
Syntax HOPControl:RFANalyzer:DELete
HOPControl:RFGenerator:DELete
Options Not applicable.
RFANalyzer or RFGenerator:FREQuency
Description Set/queries the RF ANalyzer or RF Generator hop FREQuency entry at HOPC:ADDRess.
Default GPIB unit is HZ.
Default display unit is MHZ.
Syntax HOPControl:RFANalyzer:FREQuency?
HOPControl:RFGenerator:FREQuency?
HOPControl:RFANalyzer:FREQuency <real> | [:FNUM]
HOPControl:RFGenerator:FREQuency <real> | [:FNUM]
Options Refer to Appendix B.