User`s guide
20-5
Hop Control Subsystem
CTENd
CTENd
Description This Clear To ENd command replaces the RF ANalyzer hop frequency entry and the RF
Generator hop frequency entry at HOPC:ADDRess, and all hop frequency entries after
them, with 0 MHz.
For example; the hop frequency entry at HOPC:ADDRess, HOPC:ADDRess + 1, ... up to
entry 2047 are replaced with 0 MHz.
HOPC:ADDR:NEXT values are set to ADDRess+1 modulo 2048 starting at ADDRess.
Syntax HOPControl:CTENd
Options Not applicable.
DELete
Description This DELetes the RF ANalyzer hop frequency entry and the RF Generator hop frequency
entry at HOPC:ADDRess. All other RF ANalyzer and RF Generator hop frequency entries
move down by one address. Entry 2047 in the hop table is replaced with 0 MHz.
Syntax HOPControl:DELete
Options Not applicable.
INSert
Description This INSerts an entry of 0 MHz into the RF ANalyzer hop table and the RF Generator hop
table. All other entries move down one address. Entry 2047 in the RF ANalyzer hop table
and the RF Generator hop table is lost.
Syntax HOPControl:INSert
Options Not applicable.