DSP Analyzer Commands (Measure Subsystem)
[:AMPLitude]:NPFLatness
[:AMPLitude]:NPFLatness
Description Queries the Negative Peak FLatness measurement result. This is the most negative
amplitude in dB relative to the average power over the useful bits in the measured burst.
GPIB unit is dB.
Display unit is dB.
Syntax MEASure:DSPanalyzer[:AMPLitude]:NPFLatness?
MEASure:DSPanalyzer[:AMPLitude]:NPFLatness[:MM] | [:AVG]
Options Refer to Appendices D and F.
[:AMPLitude]:PPFLatness
Description Queries the Positive Peak FLatness measurement result. This is the most positive
amplitude in dB relative to the average power over the useful bits in the measured burst.
GPIB unit is dB.
Display unit is dB.
Syntax MEASure:DSPanalyzer[:AMPLitude]:PPFLatness?
MEASure:DSPanalyzer[:AMPLitude]:PPFLatness[:MM] | [:AVG]
Options Refer to Appendices D and F.
[:AMPLitude]:PTCPower
Description Queries the Peak Transmitter Carrier Power measurement result. This is the average power
over the useful bits in the measured burst.
GPIB unit is dBm, W;
default unit is dBm.
Display unit is dBm, V, mV, uV, dBuV, W;
default unit is dBm.
Syntax MEASure:DSPanalyzer[:AMPLitude]:PTCPower?
MEASure:DSPanalyzer[:AMPLitude]:PTCPower[:MM] | [:AVG]
Options Refer to Appendices D and F.