User`s guide

15-6
DSP Analyzer Commands (Measure Subsystem)
[:AMPLitude]:MARKer:TIME:FALL
[:AMPLitude]:MARKer:TIME:FALL
Description Sets/queries the MARKer FALL trace TIME attributes.
Queries the MARKer FALL trace TIME which is the marker’s position relative to the last
bit in the measured burst. This value is a function of the fall trace marker position set or
queried by DSP:AMPL:MARK:POS:FALL.
IMPORTANT: The user MUST be on the Amplitude Fall screen to query this result
(DISP:DSP:VIEW ’AMPL FALL’).
GPIB units are seconds (S), bit periods (T);
default unit is seconds (S).
Display units are US (micro-second), T (bit periods);
default unit is US (micro-second).
Syntax MEASure:DSPanalyzer[:AMPLitude]:MARKer:TIME:FALL?
MEASure:DSPanalyzer[:AMPLitude]:MARKer:TIME:FALL[:MM] | [:AVG]
Options Refer to Appendices D and F.
[:AMPLitude]:MARKer:TIME:MID
Description Sets/queries the MARKer MID trace TIME attributes.
Queries the MARKer MID trace TIME which is the marker’s position relative to bit zero
in the measured burst. This value is a function of the mid trace marker position set or
queried by DSP:AMPL:MARK:POS:RISE.
IMPORTANT: The user MUST be on the Amplitude MID screen to query this result
(DISP:DSP:VIEW ’AMPL MID’).
GP-IB units are seconds (S), bit periods (T);
default unit is seconds (S).
Display units are US (micro-second), T (bit periods);
default unit is US (micro-second).
Syntax MEASure:DSPanalyzer[:AMPLitude]:MARKer:TIME:MID?
MEASure:DSPanalyzer[:AMPLitude]:MARKer:TIME:MID[:MM] | [:AVG]
Options Refer to Appendices D and F.