User`s guide

13-5
DISPlay Subsystem
FREeze
FREeze
Description Screen freezing prevents the Agilent 8922M/S from updating the display when running
tests. The measurement mode changes as before. This will enable tests to run more quickly.
When screen freezing is turned off, the display reverts to the last screen selected by the test
code. This is true for both manual and remote operation.
Syntax DISPlay:FREeze?
DISPlay:FREeze <string>
Options ‘ON’ | ‘OFF’
NOTE It is recommended that you select ‘ON’ from the Configure screen.
That is; DISPlay:SCReen ‘CONF’
MSYNc:BURSt:NUMBer
Description Sets/queries the MSYNc BURSt NUMBer to be displayed when the MEAS SYNC screen
is displayed.
Syntax DISPlay:MSYNc:BURSt:NUMBer?
DISPlay:MSYNc:BURSt:NUMBer <integer> | [:INUM]
Options Refer to Appendix A.
ORFSpectrum:VIEW
Description Selects/queries the Output RF Spectrum VIEW to be selected when DISPlay:SCReen
ORFSpectrum is selected.
Syntax DISPlay:ORFSpectrum:VIEW?
DISPlay:ORFSpectrum:VIEW <string>
Options ‘TRACE’ | ‘MAIN’