User`s guide
9-14
Cell Control Subsystem
CALL:SIGNaling
CALL:SIGNaling
Description Selects / queries the amount of signaling performed by the Agilent 8922M/S.
Syntax CELL:CALL:SIGNaling?
CELL:CALL:SIGNaling <string>
Options ’NORMAL’ | ’LIMITED’
Where;
• NORMAL signaling mode uses all the normal GSM messages to change the channel
configuration.
• LIMITED specifies that the Agilent 8922M/S should perform an operation with a
limited amount of signaling. The user can therefore achieve the ’force TCH’ capability
by merely pressing the ORG CALL front-panel hardkey.
CALL:TCH:ARFCn
Description Selects/queries the current traffic channel ARFCn for the current call.
Syntax CELL:CALL:TCH:ARFCn?
CELL:CALL:TCH:ARFCn <integer> | [:INUM]
Options Refer to Appendix A.
CALL:TCH:TSLot
Description Selects/queries the current traffic channel timeslot for the current call.
Syntax CELL:CALL:TCH:TSLot?
CELL:CALL:TCH:TSLot <integer> | [:INUM]
Options Refer to Appendix A.
CALL:STATus:CALLer
Description Returns ’BS’, ’MS’, or ’--’. Indicates who originated the call in progress. ’--’ indicates that
the Call STatus is inactive.
Syntax CELL:CALL:STATus:CALLer?
Options Not Applicable.