User`s guide
9-9
Cell Control Subsystem
AUDio:LOOPback:OFF
AUDio:LOOPback:OFF
Description Turns MS Loopback OFF.
Syntax CELL:AUDio:LOOPback:OFF
Options Not Applicable.
AUDio:LOOPback:NOFE
Description Turns MS Loopback on with NO Frame Erasure.
Syntax CELL:AUDio:LOOPback:NOFE
Options Not Applicable
AUDio:SPEech:CONFigure
Description Selects/queries the speech configuration.
NOTE This has couplings with DC AM.
Syntax CELL:AUDio:SPEech:CONFigure?
CELL:AUDio:SPEech:CONFigure <string>
Options ’NONE’ | ’UNCOND’ | ’COND’ | ’ECHO’ | ’PRBS’
Where;
• NONE means that speech (hardware) is deactivated.
• UNCOND means that speech will be unconditioned (that is not amplifiable).
• COND means that speech will be conditioned (i.e., amplifiable).
• ECHO means that speech (hardware) will be put into an ”echo” mode, where we will
echo back to the MS whatever we received, with a settable echo delay (time).
• PRBS means that speech (hardware) is outputting a pseudo random binary sequence.
NOTE NONE, UNCOND, COND are not available in the Agilent 8922S.