User`s guide

9-9
Cell Control Subsystem
AUDio:LOOPback:OFF
AUDio:LOOPback:OFF
Description Turns MS Loopback OFF.
Syntax CELL:AUDio:LOOPback:OFF
Options Not Applicable.
AUDio:LOOPback:NOFE
Description Turns MS Loopback on with NO Frame Erasure.
Syntax CELL:AUDio:LOOPback:NOFE
Options Not Applicable
AUDio:SPEech:CONFigure
Description Selects/queries the speech configuration.
NOTE This has couplings with DC AM.
Syntax CELL:AUDio:SPEech:CONFigure?
CELL:AUDio:SPEech:CONFigure <string>
Options ’NONE’ | ’UNCOND’ | ’COND’ | ’ECHO’ | ’PRBS’
Where;
NONE means that speech (hardware) is deactivated.
UNCOND means that speech will be unconditioned (that is not amplifiable).
COND means that speech will be conditioned (i.e., amplifiable).
ECHO means that speech (hardware) will be put into an ”echo” mode, where we will
echo back to the MS whatever we received, with a settable echo delay (time).
PRBS means that speech (hardware) is outputting a pseudo random binary sequence.
NOTE NONE, UNCOND, COND are not available in the Agilent 8922S.