User`s guide

Chapter 5 5-3
Data Processing Chain
Data Arrays
Data Arrays
Figure 5-1 shows the different kinds of data available within the instrument:
pre-raw measured data
raw measured data
error-corrected data
formatted data
trace memory
calibration coefficients
Trace memory can be directly output to a controller with OUTPMEMO;, but it cannot be
directly transmitted back.
Figure 5-1 The Data-Processing Chain For Measurement Outputs