Specifications
5-52
Hardkey/Softkey Reference
Analyzer Functions
selects ripple limit line trace on the display color
modification.
selects the ripple limit line choice. This selection leads to
menus used to define ripple limits or specifications with
which to compare a test device. Refer to the “Using Ripple
Limits to Test a Device” section in the “Making
Measurements” chapter of the user’s guide.
resets the seconds counter to zero in real-time clock.
is located in the ECal confidence check menu. This softkey
activates and displays the ECal confidence state.
presents the S-parameter menu, which is used to define
the input ports and test set direction for S-parameter
measurements.
provides a measurement calibration for reflection-only.
Measurements of one-port devices or properly terminated
two-port devices, at port 1 of an S-parameter test set. (ES
only)
is used to enter the standard numbers for the first class
required for an S
11
1-port calibration. (For default cal kits,
this is the open.)
is used to enter the standard numbers for the second class
required for an S
11
1-port calibration. (For default cal kits,
this is the short.)
is used to enter the standard numbers for the third class
required for an S
11
1-port calibration. (For default kits,
this is the load.)
measures the short circuit TRL/LRM calibration data for
PORT 1.
provides an S11 and S21 enhanced response calibration
(forward direction). Enhanced response generates a 1-port
cal for S11 and an improved calibration over the response
cal for S21 (ES only).
provides a measurement calibration for reflection-only.
Measurements of one-port devices or properly terminated
two-port devices, at port 2 of an S-parameter test set. (ES
only)
provides an S22 and S12 enhanced response calibration
(reverse direction). Enhanced response generates a 1-port
cal for S22 and an improved calibration over the response
cal for S12 (ES only).
RIPPLE LIM LINES
RIPPLE LIMIT
ROUND SECONDS
SET CONF STANDARD
S PARAMETERS
S11 1-PORT
S11A
S11B
S11C
S11 REFL SHORT
S11/S21 ENH. RESP.
S22 1-PORT
S22/S12 ENH. RESP.