Reference Guide Agilent Technologies 8753ES and 8753ET Network Analyzers Part Number 08753-90473 Printed in USA June 2002 Supersedes February 2001 © Copyright 1999–2002 Agilent Technologies, Inc.
Notice The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
How to Use This Guide This guide uses the following conventions: Front-Panel Key This represents a key physically located on the instrument. SOFTKEY This represents a “softkey,” a key whose label is determined by the instrument’s firmware. Screen Text This represents text displayed on the instrument’s screen.
Documentation Map The Installation and Quick Start Guide provides procedures for installing, configuring, and verifying the operation of the analyzer. It also will help you familiarize yourself with the basic operation of the analyzer. The User’s Guide shows how to make measurements, explains commonly-used features, and tells you how to get the most performance from your analyzer. The Reference Guide provides reference information, such as specifications, menu maps, and key definitions.
Contents 1. 8753ES Specifications and Characteristics Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Corrected System Performance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Instrument Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10 Uncorrected Port Performance . . . . . . . . .
Contents Option 1CP, Rack Mount Flange Kit With Handles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4 Service and Support Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4 Accessories Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5 Measurement Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents 10. Determining System Measurement Uncertainties Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-2 Sources of Measurement Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-3 Sources of Systematic Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-3 Sources of Random Errors . . . . . . . . . . . . . .
Contents viii
1 8753ES Specifications and Characteristics 1-1
8753ES Specifications and Characteristics Definitions Definitions All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 1/2 hour after the instrument has been turned on. Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Characteristic (char.
8753ES Specifications and Characteristics Corrected System Performance Corrected System Performance The specifications in this section apply for measurements made using 10 Hz IF bandwidth, no averaging, and at an environmental temperature of 25 ±5 °C, with less than 1 °C deviation from the calibration temperature. Assumes that an isolation calibration was performed with an averaging factor of 16.
8753ES Specifications and Characteristics Corrected System Performance Table 1-2 7-mm Device Connector Type Network Analyzer: 8753ES, Standard and Option 006 Calibration Kit: 85031B (7-mm, 50 Ω) Cables: 11857D Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 25 ± 5 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Typical Description Specification 30 to 300 kHz 300 kHz to 1.3 GHz 1.
8753ES Specifications and Characteristics Corrected System Performance Table 1-3 Type-N (50 Ω) Device Connector Type Network Analyzer: 8753ES, Standard and Option 006 Calibration Kit: 85032B/E (Type-N, 50 Ω) Cables: 11857D Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 25 ± 5 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Typical Description Specification 30 to 300 kHz 300 kHz to 1.3 GHz 1.
8753ES Specifications and Characteristics Corrected System Performance Table 1-4 3.5-mm Device Connector Type Network Analyzer: 8753ES, Standard and Option 006 Calibration Kit: 85033D (3.5-mm, 50 Ω) Cables: 11857D Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 25 ± 5°C with < 1°C deviation from cal temp, Isol cal with avg = 8 Typical Description Specification 30 to 300 kHz 300 kHz to 1.3 GHz 1.
8753ES Specifications and Characteristics Corrected System Performance Table 1-5 7/16-mm Device Connector Type Network Analyzer: 8753ES, Standard and Option 006 Calibration Kit: 85038A (7/16-mm, 50 Ω) Cables: 11857D Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 25 ± 5°C with < 1°C deviation from cal temp, Isol cal with avg = 8 Typical Description Specification 30 to 300 kHz 300 kHz to 1.3 GHz 1.
8753ES Specifications and Characteristics Corrected System Performance Table 1-6 Type-N (75Ω) Device Connector Type Network Analyzer: 8753ES, Option 075 Calibration Kit: 85036B/E (Type-N, 75Ω) Cables: 11857B Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 25 ± 5°C with < 1°C deviation from cal temp, Isol cal with avg = 8 Typical Description Specification 30 to 300 kHz 300 kHz to 1.3 GHz 1.
8753ES Specifications and Characteristics Corrected System Performance Table 1-7 Type-F (75 Ω) Device Connector Type Network Analyzer: 8753ES, Option 075 Calibration Kit: 85039B (Type-F, 75 Ω) Cables: 11857B Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 25 ± 5 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Typical Description Specification 30 to 300 kHz 300 kHz to 1.3 GHz 1.
8753ES Specifications and Characteristics Instrument Specifications Instrument Specifications Uncorrected Port Performance Table 1-8 7-mm Device Connector Type 8753ES (7-mm, 50 Ω), Standard and Option 006 Specification 30 kHz to 50 kHz 50 kHz to 300 kHz 300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 to 6 GHz (Opt 006 only) Directivitya (dB) 15, typ. 20, typ. 35 30 25 Source Matcha (dB) 10, typ. 18, typ. 16 16 14 Load Matcha (dB) 10, typ. 18, typ. 18 16 14 Reflection Trackinga (dB) ±2.5, typ.
8753ES Specifications and Characteristics Instrument Specifications Table 1-9 Type-N (75 Ω) Device Connector Type 8753ES (Type-N, 75Ω), Option 075 Description Specification 30 kHz to 50 kHz 50 kHz to 300 kHz 300 kHz to 1.3 GHz 1.3 GHz to 3 GHz 3 to 6 GHz (Opt 006 only) Directivitya (dB) 15, typ. 20, typ. 35 30 N/A Source Matcha (dB) 10, typ. 18, typ. 16 16 N/A Load Matcha (dB) 10, typ. 18, typ. 18 16 N/A Reflection Trackinga (dB) ±2.5, typ. ±2.5, typ. ±1.0 ±1.
8753ES Specifications and Characteristics Instrument Specifications Test Port Output Table 1-10 Test Port Output 8753ES Test Port Output Description Specification Supplemental Information Standard 30 kHz to 3.0 GHz 10 kHz to 3 GHz, typ. Option 006 30 kHz to 6.0 GHz 10 kHz to 6 GHz, typ. Frequency Range Resolution 1 Hz Stability Standard ±7.5 ppm, 0˚ to 55˚C, typ. ±3 ppm/year, typ. Option 1D5 0.05 ppm, 0˚ to 55˚C, typ. ±0.5 ppm/year, typ.
8753ES Specifications and Characteristics Instrument Specifications Table 1-11 Test Port Output 8753ES Test Port Output Description Specification Supplemental Information Output Powera (above 300 kHz) Power Resolution 0.01 dB Attenuator Switch Points: a. Source output performance on port 1 only. Port 2 output performance is a characteristic.
8753ES Specifications and Characteristics Instrument Specifications Table 1-12 Test Port Output 8753ES Test Port Output Description Specification Supplemental Information Output Powera (above 300 kHz) Linearityb Standard −15 to +5 dBm ±0.2 dB relative to 0 dBm output level +5 to +10 dBm ±0.5 dB relative to 0 dBm output level −15 to +5 dBm ±0.2 dB relative to 0 dBm output level +5 to +8 dBm ±0.5 dB relative to 0 dBm output level Option 014/Option 075 Impedance Standard 50 Ω, nom.
8753ES Specifications and Characteristics Instrument Specifications Table 1-13 Test Port Output 8753ES Test Port Output Description Specification Supplemental Information Signal Puritya 2nd Harmonic at maximum output power 16 MHz to 1.5 GHz (source freq.) 16 MHz to 3 GHz (Option 006, source freq.) < −25 dBc (Option 002 only) < −25 dBc, char. (non-Option 002) at 10 dB below maximum output power < −40 dBc, typ. at 20 dB below maximum output power < −50 dBc, typ.
8753ES Specifications and Characteristics Instrument Specifications Test Port Input Table 1-14 Test Port Input 8753ES Test Port Input Description Specification Supplemental Information Standard 30 kHz to 3.0 GHz 10 kHz to 3 GHz, typ. Option 006 30 kHz to 6.0 GHz 10 kHz to 6 GHz, typ. Frequency Range Frequency Response, Non-Ratio Measurement (A or B) 300 kHz to 3 GHz ±1 dB at preset power level 3 GHz to 6 GHz ±2 dB at preset power level Impedance 50 Ω, nominal.
8753ES Specifications and Characteristics Instrument Specifications Table 1-15 Test Port Input 8753ES Test Port Input Description Specification Supplemental Information Internally Generated Harmonics (Option 002 Only) 2nd Harmonic at +8 dBm input level 16 MHz to 3 GHz < −15 dBc at +0 dBm input level < −30 dBc, typ. at −15 dBm input level < −45 dBc, typ. 3rd Harmonic at +8 dBm input level 16 MHz to 2 GHz < −30 dBc at +0 dBm input level < −50 dBc, typ. at −15 dBm input level < −50 dBc, typ.
8753ES Specifications and Characteristics Instrument Specifications Table 1-16 Test Port Input 8753ES Test Port Input Description Specification Supplemental Information Standard 300 kHz to 3 GHz at −25 dBm R channel power level Option 006 300 kHz to 6 GHz External Source Modea Frequency Range R Input Requirements Power Level 0 to −25 dBm, typ. R Input Spectral Purity Requirement Maximum Spurious Input < −30 dBc, typ. Residual FM < 20 kHz, typ. Settling Time Auto 500 ms, typ.
8753ES Specifications and Characteristics Instrument Specifications Table 1-17 Test Port Input 8753ES Test Port Input Description Specification System Bandwidths 3000 Hz 10 Hz 300 kHz to 3GHz < 0.006 dB rms < 0.001 dB rms 3 GHz to 6 GHz < 0.010 dB rms < 0.002 dB rms 300 kHz to 3GHz < 0.038° rms < 0.006° rms 3 GHz to 6 GHz < 0.070° rms < 0.012° rms Trace Noisea Magnitude Phase a.
8753ES Specifications and Characteristics Instrument Specifications Table 1-19 Test Port Input 8753ES Test Port Input Dynamic Accuracy (Specification) For input ports 1 and 2, accuracy of the test port input power reading relative to the reference input power level. • Inputs: testport 1 and 2 • For test port powers > −50 dBm and < 0 dBm, magnitude dynamic accuracy is 0.02 dB + 0.001 dB/dB from the reference power, phase dynamic accuracy is 0.132 deg + 0.0066 deg/dB from the reference power.
8753ES Specifications and Characteristics Instrument Specifications Table 1-20 Test Port Input 8753ES R-Channel Input Dynamic Accuracy (Typical) Accuracy of the R-channel power reading relative to the R-channel reference power level. This table applies when the calibrated power level and the measurement power level are not the same. Inputs: • R-channel • For Option 075 and 014, for test port powers up to the maximum source power.
8753ES Specifications and Characteristics Instrument Specifications General Information Table 1-21 General Information 8753ES General Information Description Specification Display Range Magnitude ±200 dB (at 20 dB/div), max Phase ±180°, max Polar 10 pico units, min 1000 units, max Display Resolution Magnitude 0.001 dB/div, min Phase 0.01°/div, min Reference Value Range Magnitude ±500 dB, max Phase ±360 °, max Reference Level Resolution Magnitude 0.001 dB, min Phase 0.
8753ES Specifications and Characteristics Instrument Specifications Table 1-22 General Information 8753ES General Information Description Specification Supplemental Information Group Delaya Aperture (selectable) (frequency span)/(number of points −1) Maximum Aperture 20% of frequency span Range 1/2 × (1/minimum aperture) with smoothing enabled Maximum Delay Limited to measuring no more than 180° of phase change within the minimum aperture.) Accuracy See graph. Char.
8753ES Specifications and Characteristics Instrument Specifications Table 1-23 General Information 8753ES General Information Description Supplemental Information System Bandwidths IF bandwidth settings 6000 Hz, nom. 3700 Hz, nom. 3000 Hz, nom. 1000 Hz nom. 300 Hz, nom. 100 Hz, nom. 30 Hz, nom. 10 Hz, nom. Rear Panel External Auxiliary Input Connector Female BNC Range ±10 V, typ. External Trigger Triggers on a positive or negative TTL transition or contact closure to ground. Damage Level < −0.
8753ES Specifications and Characteristics Instrument Specifications Table 1-24 General Information 8753ES General Information Description Specification Supplemental Information Rear Panel Test Port Bias Input Maximum voltage ±30 Vdc Maximum current (no degradation in RF specifications) ±200 mA Maximum current ±1 A External Reference In Input Frequency 1, 2, 5, and 10 MHz ±200 Hz at 10 MHz Input Power −10 dBm to +20 dBm, typ. Input Impedance 50 Ω, nom.
8753ES Specifications and Characteristics Instrument Specifications Table 1-25 General Information 8753ES General Information Description Specification Front Panel Display Pixel Integrity Red, Green, or Blue Pixels Red, green, or blue “stuck on” pixels may appear against a black background.
8753ES Specifications and Characteristics Instrument Specifications Table 1-26 General Information 8753ES General Information Description Specification Supplemental Information General Environmental RFI/EMI Susceptibility Defined by CISPR Pub. 11 and FCC Class B standards. ESD Minimize using static-safe work procedures and an antistatic bench mat (part number 9300-0797). Dust Minimize for optimum reliability.
8753ES Specifications and Characteristics Instrument Specifications Speed Parameters Table 1-27 8753ES Measurement and Data Transfer Speed Performance Typical Time for Completion (ms) Description Number of Points 51 201 401 1601 Typical Time for Completion (in ms), Center 1 GHz, Span 10 MHz, IFBW=6000 Uncorrected 32 70 121 423 1-port and Enhanced Response cala 35 71 1271 440 2-port calb 62 139 240 848 Typical Time for Completion (in ms), Start 30 kHz, Stop 3 GHz, IFBW=6000 Uncorrected
8753ES Specifications and Characteristics Instrument Specifications Table 1-28 8753ES Recall and Sweep Speed Performance Total Time, typical (secs) Operations Channel Points Recall-Only Time, typical (secs) Raw Offset Blank Off Blank On Blank Off Blank On Error Correction ON Recall and Sweep Single Chan. 201 On 0.389 0.260 0.250 0.126 Recall and Sweep Single Chan. 201 Off 0.340 0.210 0.201 0.077 Sweep only (no Recall) Single Chan. 201 N/A 0.139 0.
8753ES Specifications and Characteristics Instrument Specifications Table 1-29 Sweep Time vs. IF Bandwidth IF Bandwidth Typical Sweep Time (seconds) a 6000 0.070 3700 0.095 3000 0.121 1000 0.248 300 0.704 100 2.022 30 6.987 10 21.365 a. Preset condition, CF = 1 GHz, Span = 100 MHz; includes retrace time, 201 points. Table 1-30 Sweep Time vs. Number of Points Number of Points Typical Sweep Time (seconds) a 51 0.039 101 0.057 201 0.095 401 0.171 801 0.323 1601 0.625 a.
8753ES Specifications and Characteristics Instrument Specifications Power Meter Calibration Accuracy Table 1-31 Power Meter Calibration Sweep Speed and Accuracy Power Desired at Test Port +5 dBm −15 dBm −30 dBm Number of Readings Sweep Time Setting (seconds) a Characteristic Accuracy (dB) b 1 33 ±0.7 2 64 ±0.2 3 95 ±0.1 1 48 ±0.7 2 92 ±0.2 3 123 ±0.1 1 194 ±0.7 2 360 ±0.2 3 447 ±0.1 a.
8753ES Specifications and Characteristics Instrument Specifications 1-32
2 8753ET Specifications and Characteristics 2-1
8753ET Specifications and Characteristics Definitions Definitions All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 1/2 hour after the instrument has been turned on. Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Characteristic (char.
8753ET Specifications and Characteristics Corrected System Performance Corrected System Performance The specifications in this section apply for measurements made using 10 Hz IF bandwidth, no averaging, and at an environmental temperature of 25 ±5 °C, with less than 1 °C deviation from the calibration temperature. Assumes that an isolation calibration was performed with an averaging factor of 16.
8753ET Specifications and Characteristics Corrected System Performance Table 2-2 Type-N (50 Ω) Device Connector Type Network Analyzer: 8753ET Standard or Option 004 Attenuator Calibration Kit: 85032B/E (Type-N, 50 Ω) Cables: 8120-5639 Calibration: See Below IF BW = 10 Hz, Avg = off, Temp = 25 ± 5 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Specification Description 300 kHz to 1.3 GHz 1.
8753ET Specifications and Characteristics Corrected System Performance Transmissiona Uncertainty: Enhanced Response Calibration (Specification) Transmission Uncertainty: Response Calibration (Specification) Reflection Uncertainty: One-Port Calibration (Specification) a. Option 004 may degrade transmission source match as much as 2 dB, resulting in up to 0.05 dB additional uncertainty in transmission uncertainty.
8753ET Specifications and Characteristics Instrument Specifications Instrument Specifications Uncorrected Port Performance Table 2-3 Type-N (50 Ω) Device Connector Type 8753ET (Type-N, 50 Ω) Specification Description 300 kHz to 1.3 GHz 1.3 to 3 GHz 3 to 6 GHz (Opt 006 only) 30 24 19 Standard (dB) 25 20 14 Option 004 (dB) 23 18 14 24 19 16 Reflection Trackinga (dB) ±1.0 ±1.0 ±2.0 Transmission Trackinga (dB) ±1.5 ±1.5 ±2.5 Magnitude (db) .002/°C, typ. .002/°C, typ. .004/°C, typ.
8753ET Specifications and Characteristics Instrument Specifications Test Port Output Table 2-4 Test Port Output 8753ET Test Port Output Description Specification Supplemental Information Standard 300 kHz to 3.0 GHz 10 kHz to 3 GHz, typ. Option 006 300 kHz to 6.0 GHz 10 kHz to 6 GHz, typ. Frequency Range Resolution 1 Hz Stability Standard ±7.5 ppm, 0° to 55 °C, typ. ±3 ppm/year, typ. Option 1D5 ±0.05 ppm, 0° to 55 °C, typ. ±0.5 ppm/year, typ.
8753ET Specifications and Characteristics Instrument Specifications Table 2-5 Test Port Output 8753ET Test Port Output Description Specification Supplemental Information Output Powera (above 300 kHz) Power Resolution 0.01 dB Attenuator Switch Points (Option 004 Only): a. Source output performance on port 1 only. Port 2 output performance is a characteristic.
8753ET Specifications and Characteristics Instrument Specifications Table 2-6 Test Port Output 8753ET Test Port Output Description Specification Supplemental Information Output Power (above 300 kHz) Linearitya Standard −20 to −15 dBm ±0.5 dB relative to −5 dBm output level −15 to 0 dBm ±0.2 dB relative to −5 dBm output level ±0.5 dB relative to −5 dBm output level −15 to +5 dBm ±0.2 dB relative to 0 dBm output level +5 to +10 dBm ±0.
8753ET Specifications and Characteristics Instrument Specifications Table 2-7 Test Port Output 8753ET Test Port Output Description Specification Supplemental Information Signal Purity 2nd Harmonic 16 MHz to 1.5 GHz (source frequency) 16 MHz to 3 GHz (Option 006, source frequency) at the maximum output power < −25 dBc, char. (Option 002) at 10 dB below maximum output power < −40 dBc, typ. at 20 dB below maximum output power < −50 dBc, typ.
8753ET Specifications and Characteristics Instrument Specifications Test Port Input Table 2-8 Test Port Input 8753ET Test Port Input Description Specification Supplemental Information Standard 300 kHz to 3.0 GHz 10 kHz to 3 GHz, typ. Option 006 300 kHz to 6.0 GHz 10 kHz to 6 GHz, typ. Frequency Range Frequency Response (Transmission) 300 kHz to 3 GHz ±1 dB at preset power level 3 GHz to 6 GHz ±2 dB at preset power level Impedance 50 Ω, nominal.
8753ET Specifications and Characteristics Instrument Specifications Table 2-9 Test Port Input 8753ET Test Port Input Description Specification Supplemental Information Internally Generated Harmonics (Option 002 Only) 2nd Harmonic 16 MHz to 3 GHz at 0 dBm input level < −15 dBc, char. at −10 dBm input level < −30 dBc, typ. at −25 dBm input level < −45 dBc, typ. 3rd Harmonic 16 MHz to 2 GHz at 0 dBm input level < −30 dBc, char. at −10 dBm input level < −50 dBc, typ.
8753ET Specifications and Characteristics Instrument Specifications Table 2-10 Test Port Input 8753ET Test Port Input Description Specification Supplemental Information Frequency Offset Operationa Frequency Range Standard 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz R Channel Input Requirements 300 kHz to 3 GHz 0 to −35 dBm 3 GHz to 6 GHz 0 to −30 dBm LO Spectral Purity and Accuracy Maximum Spurious Input < −25 dBc, typ. Residual FM < 20 kHz, typ.
8753ET Specifications and Characteristics Instrument Specifications Table 2-11 Test Port Input 8753ET Test Port Input Description Specification System Bandwidths 3000 Hz 10 Hz 300 kHz to 3GHz < 0.006 dB rms < 0.001 dB rms 3 GHz to 6 GHz < 0.010 dB rms < 0.002 dB rms 300 kHz to 3GHz < 0.038° rms < 0.006° rms 3 GHz to 6 GHz < 0.070° rms < 0.012° rms Trace Noisea Magnitude Phase a. Trace noise is defined for a ratio measurement.
8753ET Specifications and Characteristics Instrument Specifications Table 2-13 Test Port Input 8753ET Test Port Input Dynamic Accuracy (Characteristic) For the transmission port, accuracy of the test port input power reading relative to the reference input power level. • Input: transmission port • For test port powers > −60 dBm and < −10 dBm, magnitude dynamic accuracy is 0.02 dB + 0.001 dB/dB from the reference power, phase dynamic accuracy is 0.132 deg + 0.0066 deg/dB from the reference power.
8753ET Specifications and Characteristics Instrument Specifications Table 2-14 Test Port Input 8753ET Test Port Input Dynamic Accuracy (Characteristic) For the reflection port, accuracy of the test port input power reading relative to the reference input power level. • Input: reflection port • For test port powers > −50 dBm and < 0 dBm, magnitude dynamic accuracy is 0.02 dB + 0.001 dB/dB from the reference power, phase dynamic accuracy is 0.132 deg + 0.0066 deg/dB from the reference power.
8753ET Specifications and Characteristics Instrument Specifications Table 2-15 Test Port Input 8753ET R-Channel Input Dynamic Accuracy (Typical) Accuracy of the R-channel power reading relative to the R-channel reference power level. This chart applies when the calibrated power level and the measurement power level are not the same. • Inputs: R-channel 300 KHz to 3 GHz 3 to 6 GHz Magnitude Dynamic Accuracy for Test Port Powers > −50 dBm and < 0 dBm .02 dB + .001 dB/dB from the reference power .
8753ET Specifications and Characteristics Instrument Specifications General Information Table 2-16 General Information 8753ET General Information Description Specification Display Range Magnitude ± 200 dB (at 20 dB/div), max Phase ± 180°, max Polar 10 pico units, min 1000 units, max Display Resolution Magnitude 0.001 dB/div, min Phase 0.1°/div, min Reference Level Range Magnitude ± 500 dB, max Phase ± 360°, max Reference Level Resolution Magnitude 0.001 dB, min Phase 0.
8753ET Specifications and Characteristics Instrument Specifications Table 2-17 General Information 8753ET General Information Description Specification Supplemental Information Group Delaya Aperture (selectable) (frequency span)/(number of points − 1) Maximum Aperture 20% of frequency span Range 1/2 × (1/minimum aperture) Maximum Delay Limited to measuring no more than 180° of phase change within the minimum aperture. Accuracy See graph. Char.
8753ET Specifications and Characteristics Instrument Specifications Table 2-18 General Information 8753ET General Information Description Supplemental Information System Bandwidths IF bandwidth settings 6000 Hz, nom. 3700 Hz, nom. 3000 Hz, nom. 1000 Hz nom. 300 Hz, nom. 100 Hz, nom. 30 Hz, nom. 10 Hz, nom. Rear Panel External Auxiliary Input Connector Female BNC Range 10 V, typ. External Trigger Triggers on a positive or negative TTL transition or contact closure to ground. Damage Level < −0.
8753ET Specifications and Characteristics Instrument Specifications Table 2-19 General Information 8753ET General Information Description Specification Supplemental Information Rear Panel External Reference In Input Frequency 1, 2, 5, and 10 MHz ± 200 Hz at 10 MHz Input Power −10 dBm to +20 dBm, typ. Input Impedance 50 Ω, nom. VGA Video Output 15-pin mini D-Sub; female. Drives VGA compatible monitors.
8753ET Specifications and Characteristics Instrument Specifications Table 2-20 General Information 8753ET General Information Description Specification Front Panel Display Pixel Integrity Red, Green, or Blue Pixels Red, green, or blue "stuck on" pixels may appear against a black background.
8753ET Specifications and Characteristics Instrument Specifications Table 2-21 General Information 8753ET General Information Description Specification Supplemental Information General Environmental RFI/EMI Susceptibility Defined by CISPR Pub. 11 and FCC Class B standards. ETD Minimize using static-safe work procedures and an antistatic bench mat (part number 9300-0797). Dust Minimize for optimum reliability.
8753ET Specifications and Characteristics Instrument Specifications Speed Parameters Table 2-22 Measurement and Data Transfer Speed Performance Typical Time for Completion (ms) Description Number of Points 51 201 401 1601 Typical Time for Completion (in ms), Center 1 GHz, Span 10 MHz, IFBW=6000 Uncorrected 32 70 121 423 1-port and Enhanced Response cala 35 71 127 440 2-port calb 62 139 240 848 Typical Time for Completion (in ms), Start 30 kHz, Stop 3 GHz, IFBW=6000 Uncorrected 202 270
8753ET Specifications and Characteristics Instrument Specifications Table 2-23 Recall and Sweep Speed Performance Total Time, typical (secs) Operations Channel Points Recall-Only Time, typical (secs) Raw Offset Blank Off Blank On Blank Off Blank On Error Correction ON Recall and Sweep Single Chan. 201 On 0.389 0.260 0.250 0.126 Recall and Sweep Single Chan. 201 Off 0.340 0.210 0.201 0.077 Sweep only (no Recall) Single Chan. 201 N/A 0.139 0.
8753ET Specifications and Characteristics Instrument Specifications Table 2-24 Sweep Time vs. IF Bandwidth IF Bandwidth Typical Sweep Time (seconds) a 6000 0.070 3700 0.095 3000 0.121 1000 0.248 300 0.704 100 2.022 30 6.987 10 21.365 a. Preset condition, CF = 1 GHz, Span = 100 MHz; includes retrace time, 201 points. Table 2-25 Sweep Time vs. Number of Points Number of Points Typical Sweep Time (seconds)a 51 0.039 101 0.057 201 0.095 401 0.171 801 0.323 1601 0.625 a.
8753ET Specifications and Characteristics Instrument Specifications Power Meter Calibration Accuracy Table 2-26 Power Meter Calibration Sweep Speed and Accuracy Power Desired at Test Port Number of Readings Sweep Time Setting (seconds) a Characteristic Accuracy (dB) b +5 dBm 1 33 ±0.7 2 64 ±0.2 3 95 ±0.1 1 48 ±0.7 2 92 ±0.2 3 123 ±0.1 1 194 ±0.7 2 360 ±0.2 3 447 ±0.1 −15 dBm −30 dBm a.
8753ET Specifications and Characteristics Instrument Specifications 2-28
3 Front/Rear Panel 3-1
Front/Rear Panel Front Panel Features Front Panel Features CAUTION Do not mistake the line switch for the disk eject button. See the following illustrations. If the line switch is mistakenly pushed, the instrument will be turned off, losing all settings and data that have not been saved.
Front/Rear Panel Front Panel Features Figure 3-2 8753ET Front Panel The location of the following front panel features and key function blocks is shown in Figure 3-1 and Figure 3-2. These features are described in more detail later in this chapter, and in Chapter 5 , “Hardkey/Softkey Reference,”. 1. LINE switch. This switch controls ac power to the analyzer. 1 is on, 0 is off. 2. Display. This shows the measurement data traces, measurement annotation, and softkey labels.
Front/Rear Panel Front Panel Features 9. The ENTRY block. This block includes the knob, the step keys, the number pad, and the backspace key. These allow you to enter numerical data and control the markers. You can use the numeric keypad to select digits, decimal points, and a minus sign for numerical entries. You must also select a units terminator to complete value inputs.
Front/Rear Panel Analyzer Display Analyzer Display Figure 3-3 Analyzer Display (Single Channel, Cartesian Format) The analyzer display shows various measurement information: • The grid where the analyzer plots the measurement data. • The currently selected measurement parameters. • The measurement data traces. Figure 3-3 illustrates the locations of the different information labels described below.
Front/Rear Panel Analyzer Display 2. Stimulus Stop Value. This value could be any one of the following: • The stop frequency of the source in frequency domain measurements. • The stop time in time domain measurements or CW sweeps. • The upper limit of a power sweep. When the stimulus is in center/span mode, the span is shown in this space. The stimulus values can be blanked, as described under the FREQUENCY BLANK softkey in Chapter5 ,“Hardkey/Softkey Reference.
Front/Rear Panel Analyzer Display ext Waiting for an external trigger. Ofs Frequency offset mode is on. (See the “Making Mixer Measurements” chapter in the user’s guide.) Of? Frequency offset mode error, the IF frequency is not within 10 MHz of expected frequency. LO inaccuracy is the most likely cause. (See the “Making Mixer Measurements” chapter in the user’s guide.) Gat Gating is on (time domain Option 010 only).
Front/Rear Panel Analyzer Display tsH (ES or ET Option 004) Indicates that the test set hold mode is engaged. That is, a mode of operation is selected which would cause repeated switching of the step attenuator. This hold mode may be overridden. See MEASURE RESTART or NUMBER OF GROUPS in Chapter5 ,“Hardkey/Softkey Reference.” ↑ Fast sweep indicator. This symbol is displayed in the status notation block when sweep time is ≤1.0 second. When sweep time is ≥ 1.
Front/Rear Panel Analyzer Display 12. Marker Values. These are the values of the active marker, in units appropriate to the current measurement. (Refer to “Using Analyzer Display Markers” in the “Making Measurement” chapter of the user’s guide.) 13. Marker Stats, Bandwidth. These are statistical marker values that the analyzer calculates when you access the menus with the Marker Fctn key. (Refer to “Using Analyzer Display Markers” in the “Making Measurements” chapter of the user’s guide.
Front/Rear Panel Rear Panel Features and Connectors Rear Panel Features and Connectors Figure 3-4 8753ET/ES Rear Panel Figure 3-4 illustrates the features and connectors of the rear panel, described below. Requirements for input signals to the rear panel connectors are provided in the specifications and characteristics chapter. 1. GPIB connector. This allows you to connect the analyzer to an external controller, compatible peripherals, and other instruments for an automated system.
Front/Rear Panel Rear Panel Features and Connectors 7. Fan. This fan provides forced-air cooling for the analyzer. 8. 10 MHZ PRECISION REFERENCE OUTPUT. (Option 1D5) 9. 10 MHZ REFERENCE ADJUST. (Option 1D5) 10. EXTERNAL REFERENCE INPUT connector. This allows for a frequency reference signal input that can phase lock the analyzer to an external frequency standard for increased frequency accuracy.
Front/Rear Panel Rear Panel Features and Connectors 3-12
4 Menu Maps 4-1
Menu Maps Menu Maps Menu Maps This chapter contains menus maps for the hardkeys listed below. The figure number of these menu maps is listed next to the name of the hardkey. Fold Outs are located at the end of this chapter.
Menu Maps Menu Maps Figure 4-2 Menu Map for Copy 4-3
Menu Maps Menu Maps Figure 4-3 Menu Map for Display 4-4
Menu Maps Menu Maps Figure 4-4 Menu Map for Format 4-5
Menu Maps Menu Maps Figure 4-5 Menu Map for Local 4-6
Menu Maps Menu Maps Figure 4-6 Menu Map for Marker, Marker Fctn, and Marker Search 4-7
Menu Maps Menu Maps Figure 4-7 Menu Map for Meas (ET Models Only) Meas S-Parameter Menu Aux Input Menu REFLECTION RESOLUTION [LOW] TRANSMISSN AUX OUT on OFF ANALOG IN Aux Input Input Ports Menu OFF A/R Z:Ref l B/R Z:Trans A/B COUNTER: OFF Y: Ref I A ANALOG BUS Y:Trans FRAC N * Conversion Menu 1/S DIV FRAC N B R CONVERSION [ ] INPUT PORTS RETURN RETURN RETURN * Aux input menu appears only when the analog bus on OFF (service menu key under the system hardkey) is turned to ON.
Menu Maps Menu Maps Figure 4-8 Menu Map for Meas (ES Models Only) 4-9
Menu Maps Menu Maps Figure 4-9 Menu Map for Power and Sweep Setup (ET Only) 4-10
Menu Maps Menu Maps Figure 4-10 Menu Map for Power and Sweep Setup (ES Only) 4-11
Menu Maps Menu Maps Figure 4-11 Menu Map for Preset 4-12
Menu Maps Menu Maps Figure 4-12 Menu Map for Save/Recall 4-13
Menu Maps Menu Maps Figure 4-13 Menu Map for Scale Ref 4-14
Cal Correction Menu ECal Menu CORRECTION on OFF MODULE A b INTERPOL ON off REFLECTION 1-PORT ECal Pause Menu ECal Config Menu CONTINUE ECal MODULE A b ECal Confidence Power Meter Cal. Main Menu Pwr. Loss/Sens.
ECal Menu Correction Menu Cal MODULE A b CORRECTION on OFF INTERPOL on OFF S11 1-PORT 1 CALIBRATE MENU S22 1-PORT 1 ECal Pause Menu ECal Config Menu CONTINUE ECal MODULE A b ECal Service Menu Power Meter Cal. Main Menu Pwr. Loss/Sens. Lists Menu ECal STD [CONF] PWRMTR CAL OFF USE SENSOR A/B SET CONF STANDARD CONFIDENCE CHECK REFL STD [0] PARAMETER [S11] S11/S21 1, 2 ENH. RESP. OMIT ISOL ON off TRACE TYPE [DATA] TRACE TYPE [DATA] CAL KIT [ ] S22/S12 1, 2 ENH. RESP.
Select Seq Menu A TTL I/O Menu New / Modify Seq Menu SEQUENCE X SEQ X TTL OUT SEQUENCE 1 SEQ 1 TTL OUT HIGH PARALLEL OUT ALL SEQUENCE 2 SEQ 2 Seq TTL Out Menu TTL OUT LOW SET BIT SEQUENCE 3 SEQ 3 CLEAR BIT END SWEEP HIGH PULSE SEQUENCE 4 SEQ 4 PARALL IN BIT NUMBER END SWEEP LOW PULSE SEQUENCE 5 SEQ 5 PARALL IN IF BIT H SEQUENCE 6 SEQ 6 IF BIT L Seq Spec Func Menu Seq Dec Making Menu New / Modify Seq Menu Spec Func More Menu DECISION MAKING IF LIMIT TEST PASS SEQUENCE 1 SEQ 1 E
Set Clock Menu TIME STAMP ON off User Settings Menu PRESET SETTINGS Preset Settings Menu CAL INTERP ON off ROUND SECONDS SET MINUTES K36 MODE on OFF SET HOUR K39 MODE on OFF Ripple Test Menu Select Limits Menu Edit Ripple Limits Menu LIMIT LINE RIPL LIMIT on OFF FREQUENCY BAND RIPPLE LIMIT RIPL TEST on OFF MINIMUM FREQUENCY BANDWIDTH LIMIT RIPL VALUE [ OFF ] MAXIMUM FREQUENCY BW DISPLAY on OFF MAXIMUM RIPPLE BW MARKER on OFF DELETE N DB POINTS ADD MINIMUM BANDWIDTH CLEAR LIST MAXI
Set Clock Menu TIME STAMP ON off Preset Settings Menu User Settings Menu PRESET SETTINGS CAL INTERP ON off ROUND SECONDS SET MINUTES K36 MODE on OFF SET HOUR K39 MODE on OFF ***** Ripple Test Menu Select Limits Menu Edit Ripple Limits Menu LIMIT LINE RIPL LIMIT on OFF FREQUENCY BAND RIPPLE LIMIT RIPL TEST on OFF MINIMUM FREQUENCY BANDWIDTH LIMIT RIPL VALUE [ OFF ] MAXIMUM FREQUENCY BW DISPLAY on OFF MAXIMUM RIPPLE BW MARKER on OFF DELETE N DB POINTS ADD MINIMUM BANDWIDTH CLEAR LI
Menu Maps Menu Maps 4-24
5 Hardkey/Softkey Reference 5-1
Hardkey/Softkey Reference Key Reference Key Reference This chapter contains information on the following topics: • softkey and front-panel functions in alphabetical order (includes a brief description of each function) • cross reference of programming commands to key functions • cross reference of softkeys to front-panel access keys NOTE 5-2 The SERVICE MENU keys are not included in this chapter. Information on the SERVICE MENU keys can be found in the service guide.
Hardkey/Softkey Reference Where to Look for More Information Where to Look for More Information Additional information about many of the topics discussed in this chapter is located in the following areas: • "Making Measurements" in the user’s guide contains step-by-step procedures for making measurements or using particular functions.
Hardkey/Softkey Reference Analyzer Functions Analyzer Functions This section contains an alphabetical listing of softkey and front-panel functions, and a brief description of each function. . is used to add a decimal point to the number you are entering. − is used to add a minus sign to the number you are entering. is used to step up the current value of the active function. The analyzer defines the step size for different functions. No units terminator is required.
Hardkey/Softkey Reference Analyzer Functions ∆ MODE MENU goes to the delta marker menu, which is used to read the difference in values between the active marker and a reference marker. ∆ MODE OFF turns off the delta marker mode, so that the values displayed for the active marker are absolute values. ∆ REF = 1 establishes marker 1 as a reference. The active marker stimulus and response values are then shown relative to this delta reference.
Hardkey/Softkey Reference Analyzer Functions 2.92* 85056K selects the 85056K cal kit. 2.92mm other kits selects the 2.92 mm cal kit model. 3.5mmC 85033C selects the 85033C cal kit. 3.5mmD 85033D/E selects the 85033D or the 85033E cal kit. 4X: [1] [2]/[3] [4] sets up a four-graticule display with channel 2 in the upper right quadrant and channel 3 in the lower left quadrant.
Hardkey/Softkey Reference Analyzer Functions 2) adds a new frequency band to the Ripple Limit list which is indicated by the pointer >. The new frequency band is a duplicate of the most recently selected frequency band. ADDRESS: 8753 sets the GPIB address of the analyzer, using the entry controls. There is no physical address switch to set in the analyzer. The default GPIB address is 16. ADDRESS: CONTROLLER sets the GPIB address the analyzer will use to communicate with the external controller.
Hardkey/Softkey Reference Analyzer Functions AUTO FEED ON off turns the plotter auto feed function on or off when in the define plot menu. It turns the printer auto feed on or off when in the define print menu. AUTO SCALE brings the trace data in view on the display with one keystroke. Stimulus values are not affected, only scale and reference values. The analyzer determines the smallest possible scale factor that will put all displayed data onto 80% of the vertical graticule.
Hardkey/Softkey Reference Analyzer Functions Avg is used to access three different noise reduction techniques: sweep-to-sweep averaging, display smoothing, and variable IF bandwidth. Any or all of these can be used simultaneously. Averaging and smoothing can be set independently for each channel, and the IF bandwidth can be set independently if the stimulus is uncoupled. B measures the absolute power amplitude at input B. B/R calculates and displays the complex ratio of input B to input R.
Hardkey/Softkey Reference Analyzer Functions BW MARKER on OFF displays the cutoff frequencies of the bandwidth using markers on the data trace. BW TEST on OFF turns bandpass filter bandwidth testing on or off. When bandwidth testing is on, the analyzer locates the maximum point of the data trace and uses it as the reference from which to measure the filter’s bandwidth. Then, the analyzer determines the two cutoff frequencies of the bandpass filter.
Hardkey/Softkey Reference Analyzer Functions CAL FACTOR accepts a power sensor calibration factor % for the segment. CAL FACTOR SENSOR A brings up the segment modify menu and segment edit (calibration factor menu) which allows you to enter a power sensor's calibration factors. The calibration factor data entered in this menu will be stored for power sensor A. CAL INTERP ON off sets the preset state of interpolated error-correction on or off.
Hardkey/Softkey Reference Analyzer Functions CAL Z0: SYSTEM Z0 allows you to modify the characteristic impedance of the system for TRL/LRM calibration. CALIBRATE MENU leads to the calibration menu, which provides several accuracy enhancement procedures ranging from a simple frequency response calibration to a full two-port calibration.
Hardkey/Softkey Reference Analyzer Functions CH4 DATA [ ] brings up the printer color selection menu. The channel 4 data trace default color is blue for color prints. CH4 DATA LIMIT LN selects channel 4 data trace and limit line for display color modification. CH4 MEM selects channel 4 memory trace for display color modification. CH4 MEM [ ] brings up the printer color selection menu. The channel 2 memory trace default color is red for color prints.
Hardkey/Softkey Reference Analyzer Functions CLEAR BIT when the parallel port is configured for GPIO, 8 output bits can be controlled with this key. When this key is pressed, “TTL OUT BIT NUMBER” becomes the active function. This active function must be entered through the keypad number keys, followed by the x1 key. The bit is cleared when the x1 key is pressed. Entering numbers larger than 7 will result in bit 7 being cleared, and entering numbers lower than 0 will result in bit 0 being cleared.
Hardkey/Softkey Reference Analyzer Functions CONTINUE SEQUENCE resumes a paused sequence. CONTINUOUS located under the Sweep Setup key, is the standard sweep mode of the analyzer, in which the sweep is triggered automatically and continuously and the trace is updated with each sweep. CONVERSION [ ] brings up the conversion menu which converts the measured data to impedance (Z) or admittance (Y). When a conversion parameter has been defined, it is shown in brackets under the softkey label.
Hardkey/Softkey Reference Analyzer Functions DATA and MEMORY displays both the current data and memory traces. DATA ARRAY on OFF specifies whether or not to store the error-corrected data on disk with the instrument state. DATA/MEM divides the data by the memory, normalizing the data to the memory, and displays the result. This is useful for ratio comparison of two traces, for instance in measurements of gain or attenuation. DATA - MEM subtracts the memory from the data.
Hardkey/Softkey Reference Analyzer Functions DEFINE PRINT leads to the define print menu. This menu defines the printer mode (monochrome or color) and the auto-feed state. DEFINE STANDARD makes the standard number the active function, and brings up the define standard menus. The standard number (1 to 8) is an arbitrary reference number used to reference standards while specifying a class. DELAY selects the group delay format, with marker values given in seconds.
Hardkey/Softkey Reference Analyzer Functions DISK UNIT NUMBER specifies the number of the disk unit in the disk drive that is to be accessed in an external disk store or load routine. This is used in conjunction with the GPIB address of the disk drive, and the volume number, to gain access to a specific area on a disk. The access hierarchy is GPIB address, disk unit number, disk volume number. DISP MKRS ON off displays response and stimulus values for all markers that are turned on.
Hardkey/Softkey Reference Analyzer Functions DONE 1-PORT CAL finishes one-port calibration (after all standards are measured) and turns error correction on. DONE 2-PORT CAL finishes two-port calibration (after all standards are measured) and turns error correction on (ES only). DONE FWD ENH RESP. finishes the transmission portion of the enhanced response calibration. DONE LOADS finishes all the load standards when the cal kit defines more than one load standard.
Hardkey/Softkey Reference Analyzer Functions EACH SWEEP Power meter calibration occurs on each sweep. Each measurement point is measured by the power meter, which provides the analyzer with the actual power reading. The analyzer corrects the power level at that point. The number of measurement/correction iterations performed on each point is determined by the NUMBER OF READINGS softkey. This measurement mode sweeps slowly, especially when the measured power is low.
Hardkey/Softkey Reference Analyzer Functions END SWEEP HIGH PULSE sets the TTL output on TEST SEQ BNC or the test set interconnect to normally high with a 10 microseconds pulse high at the end of each sweep. END SWEEP LOW PULSE sets the TTL output on TEST SEQ BNC or the test set interconnect to normally low with a 10 µs pulse low at the end of each sweep. ENHANCED RESPONSE provides access to the series of menus used to perform an enhanced response calibration. ENH. REFL.
Hardkey/Softkey Reference Analyzer Functions EXTENSION INPUT B adds electrical delay to the input B reference plane for any B input measurements including S-parameters. EXTENSION PORT 1 extends the reference plane for measurements of S11, S21, and S12. EXTENSION PORT 2 extends the reference plane for measurements of S22, S12, and S21. EXTENSIONS on OFF toggles the reference plane extension mode.
Hardkey/Softkey Reference Analyzer Functions FIXED MKR STIMULUS changes the stimulus value of the fixed marker. Fixed marker stimulus values can be different for the two channels if the channel markers are uncoupled using the marker mode menu. To read absolute active marker stimulus values following a MKR ZERO operation, the stimulus value can be reset to zero. FIXED MKR VALUE changes the response value of the fixed marker. In a Cartesian format this is the y-axis value.
Hardkey/Softkey Reference Analyzer Functions FORWARD: OPEN a calibration standard type. FREQ OFFS MENU leads to the frequency offset menu (Option 089 only). FREQ OFFS on OFF switches the frequency offset mode on and off. FREQUENCY specifies the frequency of a calibration factor or loss value in the power meter cal loss/sensor lists. FREQUENCY BAND selects an existing frequency band to be reviewed, edited, or deleted. The maximum number of frequency bands is 12 (numbered 1 to 12).
Hardkey/Softkey Reference Analyzer Functions GATE on OFF (Option 010 only) turns gating on or off in time domain mode. GATE: CENTER (Option 010 only) allows you to specify the time at the center of the gate. GATE: SPAN (Option 010 only) allows you to specify the gate periods. GATE: START (Option 010 only) allows you to specify the starting time of the gate. GATE: STOP (Option 010 only) allows you to specify the stopping time of the gate. GATE SHAPE (Option 010 only) leads to the gate shape menu.
Hardkey/Softkey Reference Analyzer Functions GRATICULE selects the display graticule for color modification. HARMONIC MEAS (Option 002 only) leads to the harmonics menu. Measured harmonics cannot exceed the frequency range of the analyzer receiver. HARMONIC OFF (Option 002 only) turns off the harmonic measurement mode. HARMONIC SECOND (Option 002 only) selects measurement of the second harmonic. HARMONIC THIRD (Option 002 only) selects measurement of the third harmonic.
Hardkey/Softkey Reference Analyzer Functions IF LOOP COUNTER = 0 prompts the user to select a destination sequence position (SEQUENCE 1 through 6). When the value of the loop counter reaches zero, the sequence in the specified position will run. IF LOOP < > COUNTER 0 prompts the user to select a destination sequence position (SEQUENCE 1 through 6). When the value of the loop counter is no longer zero, the sequence in the specified position will run.
Hardkey/Softkey Reference Analyzer Functions k/m kilo/milli (103 / 10−3) KIT DONE (MODIFIED) terminates the cal kit modification process, after all standards are defined and all classes are specified. Be sure to save the kit with the SAVE USER KIT softkey, if it is to be used later. LABEL CLASS leads to the label class menu, to give the class a meaningful label for future reference during calibration. LABEL CLASS DONE finishes the label class function and returns to the modify cal kit menu.
Hardkey/Softkey Reference Analyzer Functions LIMIT TEST on OFF turns limit testing on or off. When limit testing is on, the data is compared with the defined limits at each measured point. Limit tests occur at the end of each sweep, whenever the data is updated, when formatted data is changed, and when limit testing is first turned on. Limit testing is available for both magnitude and phase values in Cartesian formats.
Hardkey/Softkey Reference Analyzer Functions LIST provides a tabular listing of all the measured data points and their current values, together with limit information if it is turned on. At the same time, the screen menu is presented, to enable hard copy listings and access new pages of the table. 30 lines of data are listed on each page, and the number of pages is determined by the number of measurement points specified in the stimulus menu.
Hardkey/Softkey Reference Analyzer Functions LO MENU leads to the LO menu. Allows you to configure the external source for frequency offset. LO SOURCE ADDRESS shows the GPIB address of the LO source. LOAD defines the standard type as a load (termination). Loads are assigned a terminal impedance equal to the system characteristic impedance Z0, but delay and loss offsets may still be added. If the load impedance is not Z0, use the arbitrary impedance standard definition.
Hardkey/Softkey Reference Analyzer Functions LOG MKR displays the logarithmic magnitude value and the phase of the active marker in Polar or Smith chart format. Magnitude values are expressed in dB and phase in degrees. This is useful as a fast method of obtaining a reading of the log magnitude value without changing to log magnitude format. LOOP COUNTER displays the current value of the loop counter and allows you to change the value of the loop counter.
Hardkey/Softkey Reference Analyzer Functions LOWER LIMIT M/µ sets the lower limit value for the start of the segment in a limit line list. If an upper limit is specified, a lower limit must also be defined. If no lower limit is required for a particular measurement, force the lower limit value out of range (for example −500 dB). mega/micro (106 / 10−6) MAN ’ L THRU on OFF allows the insertion of a different thru during the ECal correction routine.
Hardkey/Softkey Reference Analyzer Functions MARKER → DELAY adjusts the electrical delay to balance the phase of the DUT. This is performed automatically, regardless of the format and the measurement being made. Enough line length is added to or subtracted from the receiver input to compensate for the phase slope at the active marker position. This effectively flattens the phase trace around the active marker, and can be used to measure electrical length or deviation from linear phase.
Hardkey/Softkey Reference Analyzer Functions MARKER 2 turns on marker 2 and makes it the active marker. If another marker is present, that marker becomes inactive and is represented on the display as ∆. MARKER 3 turns on marker 3 and makes it the active marker. MARKER 4 turns on marker 4 and makes it the active marker. MARKER 5 turns on marker 5 and makes it the active marker.
Hardkey/Softkey Reference Analyzer Functions MAXIMUM FREQUENCY is used to: 1) define the highest frequency at which a calibration kit standard can be used during measurement calibration. In waveguide, this is normally the upper cutoff frequency of the standard. or 2) set the maximum frequency of the selected frequency band when setting up ripple test parameters. MAXIMUM RIPPLE sets the maximum ripple allowed of the selected frequency band. The maximum allowable ripple is 100 dB.
Hardkey/Softkey Reference Analyzer Functions MINIMUM FREQUENCY is used to: 1) define the lowest frequency at which a calibration kit standard can be used during measurement calibration. In waveguide, this must be the lower cutoff frequency of the standard, so that the analyzer can calculate dispersive effects correctly (see OFFSET DELAY ). or 2) set the minimum frequency of the selected frequency band when setting up ripple test parameters.
Hardkey/Softkey Reference Analyzer Functions MODULE A b selects the module to be used in the ECal operation. Select the module according to frequency range and connector type of the DUT. In some cases the Module B selection will not be used. MODULE INFO presents a text window that contains information about the selected module. MORE RANGES provides access to more power ranges. N 50Ω 85032B/E selects the 85032B/E cal kit. N 50Ω 85032F selects the 85032F cal kit.
Hardkey/Softkey Reference Analyzer Functions The possible values that can be entered for number of points are 3, 11, 26, 51, 101, 201, 401,801, and 1601. The number of points can be different for the two channels if the stimulus values are uncoupled. In list frequency sweep, the number of points displayed is the total number of frequency points for the defined list (see “Sweep Types” in the “Operating Concepts” chapter of the user’s guide).
Hardkey/Softkey Reference Analyzer Functions ONE-PATH 2-PORT leads to the series of menus used to perform a high-accuracy two-port calibration without an S-parameter test set. This calibration procedure effectively removes directivity, source match, load match, isolation, reflection tracking, and transmission tracking errors in one direction only. Isolation correction can be omitted for measurements of devices with limited dynamic range.
Hardkey/Softkey Reference Analyzer Functions PARALL IN IF BIT H while creating a sequence, this softkey inserts a command to jump to another sequence if the single input selected is in a high state. PARALL IN IF BIT L while creating a sequence, this softkey inserts a command to jump to another sequence if the single input selected is in a low state. PARALLEL sets the printer or plotter port to parallel. PARALLEL [COPY/GPIO] toggles the parallel output port between the copy and GPIO output modes.
Hardkey/Softkey Reference Analyzer Functions PHASE OFFSET adds or subtracts a phase offset that is constant with frequency (rather than linear). This is independent of MARKER → DELAY and ELECTRICAL DELAY . PHASE (Option 010 only) displays a Cartesian format of the phase portion of the data, measured in degrees. This format displays the phase shift versus frequency.
Hardkey/Softkey Reference Analyzer Functions PLTR PORT PARALLEL configures the analyzer for a plotter that has a parallel (centronics) interface. PLTR PORT SERIAL configures the analyzer for a plotter that has a serial (RS-232) interface. PLTR TYPE [PLOTTER] selects a pen plotter such as the HP 7440A, HP 7470A, HP 7475A, or HP 7550B as the plotter type.
Hardkey/Softkey Reference Analyzer Functions This feature may be used to compensate for attenuation non-linearities in either a directional coupler or a power splitter. Up to 12 segments may be entered, each with a different frequency and power loss value. POWER MTR toggles between 436A or 438A/437 . These power meters are GPIB compatible with the analyzer. The model number in the softkey label must match the power meter to be used.
Hardkey/Softkey Reference Analyzer Functions PRINT ALL MONOCHROME when displaying list values, prints the entire list in monochrome. When displaying operating parameters, prints all but the last page in monochrome. The data is sent to the printer as ASCII text rather than as raster graphics, which causes the printout to be faster. PRINT: COLOR sets the print command to default to a color printer. The printer output is always in the analyzer default color values.
Hardkey/Softkey Reference Analyzer Functions PWR RANGE AUTO man toggles the power range mode between auto and manual. Auto mode selects the power range based on the power selected. Manual mode limits power entry to within the ±6 to −12 dB selected range. PWRMTR CAL [ ] leads to the power meter calibration menu which provides two types of power meter calibration, continuous (each sweep) and single-sample (one sweep).
Hardkey/Softkey Reference Analyzer Functions Re/Im MKR when in the smith marker menu, Re/Im MKR displays the values of the active marker on a Smith chart as a real and imaginary pair. The complex data is separated into its real part and imaginary part. The first marker value given is the real part M cos θ, and the second value is the imaginary part M sin θ, where M = magnitude. When in the polar marker menu, Re/Im MKR displays the values of the active marker as a real and imaginary pair.
Hardkey/Softkey Reference Analyzer Functions RECALL KEYS MENU provides access to the recall keys menu where specific registers can be recalled. RECALL KEYS on OFF presents the recall keys menu as the initial menu when Save/Recall has been pressed. RECALL REG1 recalls the instrument state saved in register 1. RECALL REG2 recalls the instrument state saved in register 2. RECALL REG3 recalls the instrument state saved in register 3. RECALL REG4 recalls the instrument state saved in register 4.
Hardkey/Softkey Reference Analyzer Functions REFL STD [0] is located in the ECal service menu. This allows you to select different reflection standards in the ECal module. The number of reflection standards varies with ECal module type. REFLECTION 1-PORT leads to the reflection calibration menu (ET only). In the ECal menu, this softkey performs the reflection 1-port calibration. REMOVE ADAPTER completes the adapter removal procedure, removing the effects of the adapter being used.
Hardkey/Softkey Reference Analyzer Functions simple response calibration, an isolation standard is required. The standard normally used to correct for isolation is an impedance-matched load (usually 50 or 75 Ωs). Response and directivity calibration procedures for reflection and transmission measurements are provided in the following pages. RESTORE DISPLAY turns off the tabular listing and returns the measurement display to the screen.
Hardkey/Softkey Reference Analyzer Functions RIPL TEST on OFF turns ripple testing on or off. When ripple testing is on, the analyzer sets the lower ripple limit line at the lowest amplitude point within the frequency band and sets the upper limit line at the user-specified amplitude above. If the trace data remains at or below the upper limit line, that portion of the ripple test passes. If the trace data rises above the upper limit line within the frequency band, the test fails.
Hardkey/Softkey Reference Analyzer Functions RIPPLE LIM LINES selects ripple limit line trace on the display color modification. RIPPLE LIMIT selects the ripple limit line choice. This selection leads to menus used to define ripple limits or specifications with which to compare a test device. Refer to the “Using Ripple Limits to Test a Device” section in the “Making Measurements” chapter of the user’s guide. ROUND SECONDS resets the seconds counter to zero in real-time clock.
Hardkey/Softkey Reference Analyzer Functions S22A is used to enter the standard numbers for the first class required for an S22 1-port calibration. (For default cal kits, this is the open.) S22B is used to enter the standard numbers for the second class required for an S22 1-port calibration. (For default cal kits, this is the short.) S22C is used to enter the standard numbers for the third class required for an S22 1-port calibration. (For default kits, this is the load.
Hardkey/Softkey Reference Analyzer Functions SCALE PLOT [FULL] is the normal scale selection for plotting on blank paper. It includes space for all display annotations such as marker values, stimulus values, etc. The entire display fits within the user-defined boundaries of P1 and P2 on the plotter, while maintaining the exact same aspect ratio as the display.
Hardkey/Softkey Reference Analyzer Functions SEGMENT POWER enters absolute power values in the swept list table. The power values are restricted to the current power range setting. If port power is uncoupled, power applies to the currently selected port, otherwise it applies to both ports. (The list table only displays one port's power values at time due to limited display area.
Hardkey/Softkey Reference Analyzer Functions SEQUENCE 6 SEQ6 activates editing mode for the segment titled "SEQ6" (default title). SEQUENCE FILENAMING accesses a file naming menu which is used to automatically increment or decrement the name of a file that is generated by the network analyzer during a SEQUENCE. SERVICE MENU leads to a series of service and test menus described in detail in the service guide. SERVICE MODES a collection of common modes used for troubleshooting.
Hardkey/Softkey Reference Analyzer Functions SETUP B sets up two-graticule, four-channel display as described in the 4 PARAM HELP KEYS menu. SETUP C sets up single-graticule, four-channel display as described in the 4 PARAM HELP KEYS menu. SETUP D sets up four-graticule, four-channel display as described in the 4 PARAM HELP KEYS menu. Two of the graticules are in Smith chart format with the other two in log format.
Hardkey/Softkey Reference Analyzer Functions In single segment mode, selecting a measurement calibration will force the full list sweep before prompting for calibration standards. The calibration will then be valid for any single segment. If an instrument state is saved in memory with a single-segment trace, a recall will re-display that segment while also recalling the entire list. SLIDING defines the load as a sliding load.
Hardkey/Softkey Reference Analyzer Functions SOURCE PWR on OFF use this key to restore power after a power interruption. ON returns the source power to its original setting, while OFF sets the source to the minimum power level of the analyzer. SPACE inserts a space in the title. Span is used, along with the Center key, to define the frequency range of the stimulus. When the Span key is pressed, it becomes the active function.
Hardkey/Softkey Reference Analyzer Functions STD TYPE: ARBITRARY IMPEDANCE defines the standard type to be a load, but with an arbitrary impedance (different from system Z0). STD TYPE: DELAY/THRU defines the standard type as a transmission line of specified length, for calibrating transmission measurements. STD TYPE: LOAD defines the standard type as a load (termination).
Hardkey/Softkey Reference Analyzer Functions STORE SEQ TO DISK presents the store sequence to disk menu with a list of sequences that can be stored. SWEEP is used to set the frequency of the LO source to sweep. SWEEP SETUP provides access to a series of menus which are used to define and control all stimulus functions other than start, stop, center, and span. Operating parameters such as power, sweeptime, trigger condition, and number of points are accessible through this hardkey.
Hardkey/Softkey Reference Analyzer Functions TALKER/LISTENER is the mode normally used for remote programming of the analyzer. In this mode, the analyzer and all peripheral devices are controlled from the external controller. The controller can command the analyzer to talk, and the plotter or other device to listen. The analyzer and peripheral devices cannot talk directly to each other unless the computer sets up a data path between them.
Hardkey/Softkey Reference Analyzer Functions TESTSET I/O REV is used to support specialized test sets, such as a test set that measures duplexers. It allows you to set three bits (D1, D2, and D3) to a value of 0 to 7, and outputs it as binary from the rear panel test set connector. It tracks the coupling flag, so if coupling is on, and REV channel 1 is the active channel, REV channel 2 will be set to the same value. TESTS presents the service test menu.
Hardkey/Softkey Reference Analyzer Functions If the value of the loop counter is zero, then the title number goes in the last point of memory. If the loop counter is greater than or equal to the current number of measurement points, the number is placed in the first point of memory. A data to memory command must be executed before using the title to memory command.
Hardkey/Softkey Reference Analyzer Functions When tracking is off, the target is found on the current sweep and remains at the same stimulus value regardless of changes in trace response value with subsequent sweeps. A maximum and a minimum point can be tracked simultaneously using two channels and uncoupled markers. TRANS: FWD S21 (B/R) defines the measurement as S21, the complex forward transmission coefficient (magnitude and phase) of the test device.
Hardkey/Softkey Reference Analyzer Functions UNCOUPLED allows the marker stimulus values to be controlled independently on each channel. UP CONVERTER sends the sum frequency of the RF and LO to the R channel. UPPER LIMIT sets the upper limit value for the start of the segment. If a lower limit is specified, an upper limit must also be defined. If no upper limit is required for a particular measurement, force the upper limit value out of range (for example +500 dB).
Hardkey/Softkey Reference Analyzer Functions VELOCITY FACTOR enters the velocity factor used by the analyzer to calculate equivalent electrical length in distance-to-fault measurements using the time domain option. Values entered should be less than 1. Velocity factor is the ratio of the velocity of wave propagation in a coaxial cable to the velocity of wave propagation in free space. Most cables have a relative velocity of about 0.66 the speed in free space.
Hardkey/Softkey Reference Analyzer Functions WIDTH VALUE is used to set the amplitude parameter (for example 3 dB) that defines the start and stop points for a bandwidth search. The bandwidth search feature analyzes a bandpass or band reject trace and calculates the center point, bandwidth, and Q (quality factor) for the specified bandwidth. Bandwidth units are the units of the current format.
Hardkey/Softkey Reference Analyzer Functions XMIT CNTRL [ ] toggles the PLOTTER/PRINTER serial port data transmit control mode between the Xon-Xoff protocol handshake and the DTR-DSR (data terminal ready-data set ready) hardwire handshake. Y: REFL converts reflection data to its equivalent admittance values. Y: TRANS converts transmission data to its equivalent admittance values. Z: REFL converts reflection data to its equivalent impedance values.
Hardkey/Softkey Reference Analyzer Functions 5-70
6 Error Messages 6-1
Error Messages Error Messages Error Messages This chapter contains the following information to help you interpret any error messages that may be displayed on the analyzer LCD or transmitted by the instrument over GPIB: • An alphabetical listing of all error messages, including: ❏ An explanation of the message ❏ Suggestions to help solve the problem • A numerical listing of all error messages NOTE 6-2 Some messages described in this chapter are for information only and do not indicate an error condition.
Error Messages Error Messages in Alphabetical Order Error Messages in Alphabetical Order ABORTING COPY OUTPUT Information Message This message is displayed briefly if you have pressed Local to abort a copy operation. If the message is not subsequently replaced by error message number 25, PRINT ABORTED (or PLOT ABORTED) the copy device may be hung. Press Local once more to exit the abort process and verify the status of the copy device.
Error Messages Error Messages in Alphabetical Order ARGUMENT OUT OF RANGE Error Number 206 The argument for a programming command is out of the specified range. Refer to the programmer’s guide for a list of programming commands and argument ranges. ASCII: MISSING 'BEGIN' STATEMENT Error Number 193 The CITIfile you just downloaded over the GPIB or via disk was not properly organized. The analyzer is unable to read the “BEGIN” statement.
Error Messages Error Messages in Alphabetical Order BATTERY FAILED. STATE MEMORY CLEARED Error Number 183 The battery protection of the non-volatile CMOS memory has failed. The CMOS memory has been cleared. Refer to the service guide for battery replacement instructions. See Chapter 8 , “Preset State and Memory Allocation” for more information about the CMOS memory. BATTERY LOW! STORE SAVE REGS TO DISK Error Number 184 The battery protection of the non-volatile CMOS memory is in danger of failing.
Error Messages Error Messages in Alphabetical Order CANNOT FORMAT DOS DISKS ON THIS DRIVE Error Number 185 You have attempted to initialize a floppy disk to DOS format on an external disk drive that does not support writing to all 80 tracks of the double density and high density disks. The older single-sided disks had only 66 tracks and some disk drives were limited to accessing that number of tracks. To format the disk, either choose another external disk drive or use the analyzer's internal disk drive.
Error Messages Error Messages in Alphabetical Order CH1 (CH2, CH3, CH4) TARGET VALUE NOT FOUND Error Number 159 Your target value for the marker search function does not exist on the current data trace. CONTINUOUS SWITCHING NOT ALLOWED Error Number 10 Your current measurement requires different power ranges on channel 1 and channel 2. To protect the attenuator from undue mechanical wear, test set hold will be activated.
Error Messages Error Messages in Alphabetical Order CORRECTION TURNED OFF Error Number 66 Critical parameters in your current instrument state do not match the parameters for the calibration set, therefore correction has been turned off. The critical instrument state parameters are sweep type, start frequency, frequency span, and number of points. CURRENT PARAMETER NOT IN CAL SET Error Number 64 Correction is not valid for your selected measurement parameter.
Error Messages Error Messages in Alphabetical Order DISK IS WRITE PROTECTED Error Number 48 The store operation cannot write to a write-protected disk. Slide the write-protect tab over the write-protect opening in order to write data on the disk. DISK MEDIUM NOT INITIALIZED Error Number 40 You must initialize the disk before it can be used. DISK MESSAGE LENGTH ERROR Error Number 190 The analyzer and the external disk drive aren't communicating properly.
Error Messages Error Messages in Alphabetical Order ECal: CHECKING ORIENTATION Information Message This message is displayed while the network analyzer verifies that the ECal module is properly connected in the RF path. ECal: CONFIDENCE STATE Information Message This message is displayed when the confidence state is initiated. The confidence state response will be displayed. ECal FAILED Error Number 224 This is a serious ECal failure.
Error Messages Error Messages in Alphabetical Order ECal: SELECT OTHER MODULE Error Number 221 This error occurs when two ECal modules are connected. The selected ECal module may not be able to measure the selected frequency range. Check the frequency range of the other module and connect in the RF path. ECal: THRU STATE Information Message This message is displayed when the ECal module is set to the thru state.
Error Messages Error Messages in Alphabetical Order FIRST CHARACTER MUST BE A LETTER Error Number 42 The first character of a disk file title or an internal save register title must be an alpha character. FORMAT NOT VALID FOR MEASUREMENT Error Number 75 Conversion measurements (Z or Y reflection and transmission) are not valid with Smith chart and SWR formats.
Error Messages Error Messages in Alphabetical Order 8753 SOURCE PARAMETERS CHANGED Error Number 61 Some of the stimulus parameters of the instrument state have been changed, because you have turned correction on. A calibration set for the current measurement parameter was found and activated. The instrument state was updated to match the stimulus parameters of the calibration state.
Error Messages Error Messages in Alphabetical Order INSUFFICIENT MEMORY FOR PRINT/PLOT Error Number 168 There is not enough memory available for the print or plot function. Increase the available memory by changing or eliminating a memory-intensive operation such as reducing the number of points in the sweep. INSUFFICIENT MEMORY, PWR MTR CAL OFF Error Number 154 There is not enough memory space for the power meter calibration array.
Error Messages Error Messages in Alphabetical Order LOW PASS: FREQ LIMITS CHANGED Information Message The frequency domain data points must be harmonically related from dc to the stop frequency. That is, stop = n × start, where n = number of points. If this condition is not true when a low pass mode (step or impulse) is selected and transform is turned on, the analyzer resets the start and stop frequencies.
Error Messages Error Messages in Alphabetical Order NO LIMIT LINES DISPLAYED Error Number 144 You can turn limit lines on but they cannot be displayed on polar or Smith chart display formats. NO MARKER DELTA - SPAN NOT SET Error Number 15 You must turn the delta marker mode on, with at least two markers displayed, in order to use the MARKER → SPAN softkey function. NO MEMORY AVAILABLE FOR INTERPOLATION Error Number 123 You cannot perform interpolated error correction due to insufficient memory.
Error Messages Error Messages in Alphabetical Order NOT ENOUGH SPACE ON DISK FOR STORE Error Number 44 The store operation will overflow the available disk space. Insert a new disk or purge files to create free disk space. NO VALID MEMORY TRACE Error Number 54 If you are going to display or otherwise use a memory trace, you must first store a data trace to memory.
Error Messages Error Messages in Alphabetical Order OVERLOAD ON REFL PORT, POWER REDUCED (ET only) Error Number 58 See OVERLOAD ON INPUT R, POWER REDUCED (error number 57). OVERLOAD ON TRANS PORT, POWER REDUCED (ET only) Error Number 59 See OVERLOAD ON INPUT R, POWER REDUCED (error number 57). OVERLOAD ON INPUT R, POWER REDUCED Error Number 57 You have exceeded approximately +14 dBm at one of the test ports. The RF output power is automatically reduced to −85 dBm.
Error Messages Error Messages in Alphabetical Order PLOT ABORTED Error Number 27 When you press the Local key, the analyzer aborts the plot in progress. PLOTTER: not on, not connect, wrong addrs Error Number 26 The plotter does not respond to control. Verify power to the plotter, and check the GPIB connection between the analyzer and the plotter. Ensure that the plotter address recognized by the analyzer matches the GPIB address set on the plotter itself.
Error Messages Error Messages in Alphabetical Order PRINT ABORTED Error Number 25 When you press the Local key, the analyzer aborts output to the printer. print color not supported with EPSON Error Number 178 You have defined the printer type as EPSON-P2. Color print is not supported with this printer. The print will abort. PRINTER: busy Error Number 176 The parallel port printer is not accepting data. PRINTER: error Error Number 175 The parallel port printer is malfunctioning.
Error Messages Error Messages in Alphabetical Order PRINTER: power off Error Number 174 The power to the printer at the parallel port is off. PRINTER: reset in progress Information Message If the printer takes longer than a half-second to reset, this message will be displayed until printer is finished with reset.
Error Messages Error Messages in Alphabetical Order SAVE FAILED. INSUFFICIENT MEMORY Error Number 151 You cannot store an instrument state in an internal register due to insufficient memory. Increase the available memory by clearing one or more save/recall registers and pressing Preset , or by storing files to a disk. SEGMENT #n POWER OUTSIDE RANGE LIMIT Information Message The selected power range does not support the power level of one or more segments in the swept list table.
Error Messages Error Messages in Alphabetical Order SLIDES ABORTED (MEMORY REALLOCATION) Error Number 73 You cannot perform sliding load measurements due to insufficient memory. Increase the available memory by clearing one or more save/recall registers and pressing Preset , or by storing files to a disk and then deleting them from internal memory.
Error Messages Error Messages in Alphabetical Order SWEEP TRIGGER SET TO HOLD Information Message The instrument is in a hold state and is no longer sweeping. To take a new sweep, press Sweep Setup TRIGGER MENU SINGLE or CONTINUOUS . SYNTAX ERROR Error Number 33 You have improperly formatted a GPIB command. Refer to the programmer’s guide for proper command syntax. SYST CTRL OR PASS CTRL IN LOCAL MENU Error Number 36 The analyzer is in talker/listener mode.
Error Messages Error Messages in Alphabetical Order VALID ONLY FOR BILATERAL DEVICES Information Message This message is displayed when an enhanced reflection calibration is initiated. If the device tested is not a bilateral device, the enhanced reflection calibration will cause errors in the measurement results. A bilateral device has similar forward and reverse transmission characteristics. Examples of bilateral devices are passive devices (filters, attenuators, and switches).
Error Messages Error Messages in Numerical Order Error Messages in Numerical Order Error Number Error 1 OPTIONAL FUNCTION; NOT INSTALLED 2 INVALID KEY 3 CORRECTION CONSTANTS NOT STORED 4 PHASE LOCK CAL FAILED 5 NO IF FOUND: CHECK R INPUT LEVEL 6 POSSIBLE FALSE LOCK 7 NO PHASE LOCK: CHECK R INPUT LEVEL 8 PHASE LOCK LOST 9 LIST TABLE EMPTY 10 CONTINUOUS SWITCHING NOT ALLOWED 11 SWEEP TIME INCREASED 12 SWEEP TIME TOO FAST 13 AVERAGING INVALID ON NON-RATIO MEASURE 14 FUNCTION NOT V
Error Messages Error Messages in Numerical Order Error Number Error 32 WRITE ATTEMPTED WITHOUT SELECTING INPUT TYPE 33 SYNTAX ERROR 34 BLOCK INPUT ERROR 35 BLOCK INPUT LENGTH ERROR 36 SYST CTRL OR PASS CTRL IN LOCAL MENU 37 ANOTHER SYSTEM CONTROLLER ON GPIB 38 DISK: not on, not connected, wrong addrs 39 DISK HARDWARE PROBLEM 40 DISK MEDIUM NOT INITIALIZED 41 NO DISK MEDIUM IN DRIVE 42 FIRST CHARACTER MUST BE A LETTER 43 ONLY LETTERS AND NUMBERS ARE ALLOWED 44 NOT ENOUGH SPACE ON
Error Messages Error Messages in Numerical Order Error Number Error 62 NOT VALID FOR PRESENT TEST SET 63 CALIBRATION REQUIRED 64 CURRENT PARAMETER NOT IN CAL SET 65 CORRECTION AND DOMAIN RESET 66 CORRECTION TURNED OFF 67 DOMAIN RESET 68 ADDITIONAL STANDARDS NEEDED 69 NO CALIBRATION CURRENTLY IN PROGRESS 70 NO SPACE FOR NEW CAL.
Error Messages Error Messages in Numerical Order Error Number Error 133 THIS LIST FREQ INVALID IN HARM/3 GHZ RNG 141 STOP/CW FREQ + OFFSET MUST BE < 3 GHz 144 NO LIMIT LINES DISPLAYED 145 SWEEP TYPE CHANGED TO LINEAR SWEEP 148 EXTERNAL SOURCE MODE REQUIRES CW TIME 150 LOG SWEEP REQUIRES 2 OCTAVE MINIMUM SPAN 151 SAVE FAILED / INSUFFICIENT MEMORY 152 D2/D1 INVALID: CH1 CH2 NUM PTS DIFFERENT 153 SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE 154 INSUFFICIENT MEMORY, PWR MTR CAL OFF 157 SEQU
Error Messages Error Messages in Numerical Order Error Number Error 179 POWER UNLEVELED 180 DOS NAME LIMITED TO 8 CHARS + 3 CHAR EXTENSION 181 BAD FREQ FOR HARMONIC OR FREQ OFFSET 182 LIST MODE OFF: INVALID WITH LO FREQ 183 BATTERY FAILED.
Error Messages Error Messages in Numerical Order Error Number Error 212 ANALOG BUS DISABLED IN 6 KHZ IF BW 213 RANGE CAUSED POWER LVL CHANGE IN LIST 214 CORRECTION ON: AUX CHANNEL(S) RESTORED 215 CAUTION: CORRECTION OFF: AUX CHANNEL(S) DISABLED 218 CAUTION: FLOPPY DISK IS FULL 219 ECal MODULE NOT IN RF PATH 220 SELECTED MODULE OUTSIDE START-STOP FREQ RANGE 221 ECal SELECT OTHER MODULE 222 ECal MODULE NOT RESPONDING 223 ISOL AVGS < SWP AVGS 224 ECal FAILED 6-31
Error Messages Error Messages in Numerical Order 6-32
7 Options and Accessories 7-1
Options and Accessories Using This Chapter Using This Chapter This chapter contains information on the following subjects: • “Analyzer Options Available” on page 7-3 • “Accessories Available” on page 7-5 7-2
Options and Accessories Analyzer Options Available Analyzer Options Available Option 1D5, High Stability Frequency Reference Option 1D5 offers ±0.05 ppm temperature stability from 0 to 55 °C (referenced to 25 °C), and aging rate of ±0.5 ppm per year (typical). Option 002, Harmonic Mode Provides measurement of second or third harmonics of the test device's fundamental output signal. Frequency and power sweep are supported in this mode.
Options and Accessories Analyzer Options Available Option 075, 75 Ω Impedance (ES Only) Option 075 offers 75 ohm impedance bridges with type-N (f) test port connectors. Option 1CM, Rack Mount Flange Kit Without Handles Option 1CM is a rack mount kit containing a pair of flanges and the necessary hardware to mount the instrument, with handles detached, in an equipment rack with 482.6-mm (19 inches) horizontal spacing.
Options and Accessories Accessories Available Accessories Available For accessories not listed in this section, refer to the configuration guide for your analyzer or refer to the following Internet site: www.agilent.com/find/8753 Measurement Accessories Accessories are available in these connector types: 3.5-mm, 7-mm, 50 Ω type-N, 7-16, 75 Ω type-N, and type-F. A standard 8753ES or one equipped with Option 014 or Option 075 includes a built-in test set.
Options and Accessories Accessories Available • 85032F 50 Ω type-N calibration kit (30 kHz to 9 GHz) Contains a female and male open and short, female and male load standard, and a torque wrench. — Option 500 Adds four type-N to 7- mm adapter. • 85033D 3.5-mm calibration kit Contains fixed loads, one-piece open and short circuits, and 7-mm to 3.5-mm adapters for both connector sexes for use with 7-mm test-port cables. — Option 001 Deletes 7-mm to 3.5-mm adapters.
Options and Accessories Accessories Available RF electronic calibration (ECal) modules and PC software This product family provides electronic calibration (ECal) capability. With ECal, the usual calibration kit standards are replaced by one solid-state calibration module. A full two-port calibration can be done with a single connection, with reduced errors and wear on connectors.
Options and Accessories Accessories Available • 85096B RF ECal module (30 kHz to 3 GHz), 75 Ω type-N (m) to 75 Ω type-N (f) connectors — Option 00F substitutes module with two type-N female connectors. — Option 00M substitutes module with two type-N male connectors. — Option 00A adds: — type-N (m) to type-N (m) adapter — type-N (f) to type-N (f) adapter • 85099B RF ECal module (30 kHz to 3 GHz), 75 Ω type-F (m) to type-F (f) connectors — Option 00F substitutes module with two type-F female connectors.
Options and Accessories Accessories Available Verification Kit 85029B 7-mm verification kit Includes attenuators and mismatch attenuator with data on a 3.5 inch disk for use in confirming accuracy enhanced system measurement performance, traceable to national standards. Test procedure is provided in the service guide. For use with a standard 8753ES, or with systems including an 8753ES Option 011 and an 85046A, or 85047A test set.
Options and Accessories Accessories Available • 11856A 75 Ω BNC accessory kit Contains 75 Ω type-N to 75 Ω BNC adapters for both connector sexes, a BNC male short and BNC male termination. • 11906A 7-16 to 7-16 adapter kit Contains one 7-16(m) to 7-16(m) adapter, one 7-16(f) to 7-16(f) adapter, and two 7-16(m) to 7-16(f) adapters. • 11906B 7-16 to 50 Ω type-N adapter kit Contains adapters for type-N (m) to 7-16(m), type-N (m) to 7-16(f), type-N (f) to 7-16(m), and type-N (f) to 7-16(f).
Options and Accessories Accessories Available Power Sensors For more accurate control of leveled test-port power. Requires a 436A, 437B, 438B, or E4419B power meter and an GPIB cable for connection to the network analyzer. • 8481A 10 MHz to 18 GHz, type-N (m), 100 mwatt • 8481B 10 MHz to 18 GHz, type-N (m), 25 watt • 8482A 100 kHz to 4.2 GHz, type-N (m), 100 mwatt • 8482B 100 kHz to 4.2 GHZ, type-N (m), 25 watt • 8483A 100 kHz to 2 GHZ, 75 Ω type-N (m), 100 mwatt • 8485A 50 MHz to 26.5 GHz, 3.
Options and Accessories Accessories Available Keyboard Template The analyzer is designed to accept most PC-AT-compatible keyboards with a mini-DIN connector. The keyboard can be used for control or data input, such as titling files. The information found on the analyzer keyboard template (part number 08753-80220) is also listed in Table 7-2.
8 Preset State and Memory Allocation 8-1
Preset State and Memory Allocation Preset State Preset State When the Preset key is pressed, the analyzer reverts to a known state called the factory preset state. This state is defined in Table 8-1. There are subtle differences between the preset state and the power-up state. These differences are documented in Table 8-3. If power to non-volatile memory is lost, the analyzer will have certain parameters set to default settings. The affected parameters are shown in Table 8-4.
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions Preset Conditions Preset Value Analyzer Mode Analyzer Mode Network Analyzer Mode Frequency Offset Operation Off Offset Value 0 Harmonic Operation Off Stimulus Conditions Sweep Type Linear Frequency Display Mode Start/Stop Trigger Type Continuous External Trigger Off Sweep Time 87.5 ms, Auto Mode Start Frequency ES: 30 kHz. ET: 300 kHz Frequency Span ES: 2999.97 MHz ET: 2999.7 MHz Frequency Span (Opt.
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Frequency List Frequency List Empty Edit Mode Start/Stop, Number of Points Response Conditions Parameter (ES) Channel 1: S11 Channel 2: S21 Channel 3: S12 Channel 4: S22 Parameter (ET) Channel 1: Refl Channel 2: Trans Channel 3: Refl Channel 4: Trans Conversion Off Format Log Magnitude (all inputs) Display Data Color Selections Same as before Preset Dual Channel Off Act
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Smoothing Aperture 1% SPAN; Off Phase Offset 0 Degrees Electrical Delay 0 ns Scale/Division 10 dB/Division Calibration Correction Off Calibration Type None Calibration Kit (ES) 7-mm, type-N 75-ohm (Option 075) Calibration Kit (ET) Type-N 50 Ohms Enhanced Reflection Calibration Off System Z0 50 Ohms Velocity Factor 1 Extensions Off Port 1 0s Port 2 0s Input A
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Markers (coupled) Markers 1, 2, 3, 4, 5 1 GHz; All Markers Off Last Active Marker 1 Reference Marker None Marker Mode Continuous Display Markers On Delta Marker Mode Off Coupling On Marker Search Off Marker Target Value −3 dB Marker Width Value −3 dB; Off Marker Tracking Off Marker Stimulus Offset 0 Hz Marker Value Offset 0 dB Marker Aux Offset (Phase) 0 Degree
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Ripple Test Preset Value Off Bandwidth Limit Bandwidth Test Off Bandwidth Display Off Bandwidth Marker Off Time Domain Transform Off Transform Type Bandpass Start Transform −20 nanoseconds Transform Span 40 nanoseconds Gating Off Gate Shape Normal Gate Start −10 nanoseconds Gate Span 20 nanoseconds Demodulation Off Window Normal Use Memory Off System Parameters GPIB Addres
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value GPIB Address Last Active State Printer Type Last Active State Printer Port Last Active State Printer Baud Rate Last Active State Printer Handshake Last Active State Printer GPIB Address Last Active State Disk Save Configuration (Define Store) Data Array Off Raw Data Array Off Formatted Data Array Off Graphics Off Data Only Off Directory Size Defaulta Save Using B
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Plot Plot Data On Plot Memory On Plot Graticule On Plot Text On Plot Marker On Autofeed On Plot Quadrant Full Page Scale Plot Full Plot Speed Fast Pen Number: Ch1/Ch3 Data 2 Ch2/Ch4 Data 3 Ch1/Ch3 Memory 5 Ch2/Ch4 Memory 6 Ch1/Ch3 Graticule 1 Ch2/Ch4 Graticule 1 Ch1/Ch3 Text 7 Ch2/Ch4 Text 7 Ch1/Ch3 Marker 7 Ch2/Ch4 Marker 7 Line Type: Ch1/Ch3 Data
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Printer Colors Ch1/Ch3 Data Magenta Ch1/Ch3 Mem Green Ch2/Ch4 Data Blue Ch2/Ch4 Mem Red Graticule Cyan Warning Black Text Black Reference Line Black a. The directory size is calculated as 0.013% of the floppy disk size (which is ≈256) or 0.005% of the hard disk size. b. Pressing preset turns off sequencing modify (edit) mode and stops any running sequence.
Preset State and Memory Allocation Preset State Table 8-3 Power-On Conditions (versus Preset) GPIB MODE Talker/listener. SAVE REGISTERS Power meter calibration data and calibration data not associated with an instrument state are cleared. COLOR DISPLAY Default color values. SEQUENCES Sequence 1 through 5 are erased. DISK DIRECTORY Cleared.
Preset State and Memory Allocation Memory Allocation Memory Allocation The analyzer is capable of saving complete instrument states for later retrieval. It can store these instrument states into the internal memory, to the internal disk, or to an external disk.
Preset State and Memory Allocation Memory Allocation NOTE Even though calibration data is stored in non-volatile memory, if the associated instrument state is not saved, you will not be able to retrieve the calibration data after a power cycle.
Preset State and Memory Allocation Memory Allocation Determining Memory Requirements Table 8-5 shows the memory requirements of calibration arrays and memory trace arrays to help you approximate memory requirements. For example, add the following memory requirements: • a full 2-port calibration with 801 points (58 k) • the memory trace array (4.9 k) • the instrument state (6 k) The total memory requirement is 68.9 kbytes. There is sufficient memory to store 29 calibrations of this type.
Preset State and Memory Allocation Memory Allocation Table 8-5 Memory Requirements of Calibration and Memory Trace Arrays Variable Approximate Totals (Bytes) Data Length (Bytes)a 401 pts 801 pts 1 chan 1601 pts 1 chan 2 chans Calibration Arrays Response N × 6 + 52 2.5 k 5k 10 k 19 k Response and isolation N × 6 × 2 + 52 5k 10 k 19 k 38 k N × 6 × 3 + 52 7k 14 k 29 k 58 k N × 6 × 12 + 52 29 k 58 k 115 k 230 k (Nc × 2 × number channelsd) +208 1k 1.8 k 3.4 k 6.
Preset State and Memory Allocation Memory Allocation Storing Data to Disk You can use the internal disk drive or connect an external disk drive for storage of instrument states, calibration data, measurement data, and plot files. (Refer to the “Printing, Plotting, and Saving Measurement Results” chapter in the user’s guide for more information on saving measurement data and plot files.) The analyzer displays one file name per stored instrument state when you list the disk directory.
Preset State and Memory Allocation Memory Allocation Table 8-6 Suffix Character Definitions Char 1 Definition I, P Instrument statea W Four-channel instrument state G Graphics D R F Error corrected data Raw data Formatted data Char 2 Definition 1 Display graphics 0 Graphics index 1 Channel 1 2 Channel 2 3 Channel 3 4 Channel 4 1 to 4 Channel 1/3, raw arrays 1 to 4b 5 to 8 Channel 2/4, raw arrays 5 to 8 1 Channel 1 2 Channel 2 3 Channel 3 4 Channel 4 C Cal K Cal ki
Preset State and Memory Allocation Memory Allocation If correction is on at the time of an external store, the calibration set is stored to disk. (Note that inactive calibrations are not stored to disk.) When an instrument state is loaded into the analyzer from disk, the stimulus and response parameters are restored first. If correction is on for the loaded state, the analyzer will load a calibration set from disk that carries the same title as the one stored for the instrument state.
Preset State and Memory Allocation Memory Allocation Clearing Memory For security reasons, you may have the need to clear the network analyzer’s internal memory. As explained in “Types of Memory and Data Storage” on page 8-12, the internal memory consists of the 2 Mbyte non-volatile CMOS memory and the 4 Mbyte volatile memory (RAM). Clearing the internal memory clears all instrument states and measured data from the network analyzer. To clear the internal memory: 1.
Preset State and Memory Allocation Memory Allocation 8-20
9 Understanding the CITIfile Data Format 9-1
Understanding the CITIfile Data Format Using This Chapter Using This Chapter The descriptions and examples shown in this chapter demonstrate how CITIfile may be used to store and transfer both measurement information and data. The use of a single, common format will allow data to be more easily moved between instruments and computers.
Understanding the CITIfile Data Format The CITIfile Data Format The CITIfile Data Format Description and Overview CITIfile is a standardized data format, used for exchanging data between different computers and instruments. CITIfile is an abbreviation for “Common Instrumentation Transfer and Interchange file”. This standard has been a group effort between instrument designers and designers of computer-aided design programs.
Understanding the CITIfile Data Format The CITIfile Data Format A CITIfile Package A typical package is divided into two parts: The first part, the header, is made up of keywords and setup information. The second part, the data, usually consists of one or more arrays of data. Example 1 shows the basic structure of a CITIfile package: Example 1, A CITIfile Package The “header” part CITIFILE A.01.00 NAME MEMORY VAR FREQ MAG 3 DATA S RI The “data” part BEGIN -3.54545E-2, -1.38601E-3 0.23491E-3, -1.
Understanding the CITIfile Data Format The CITIfile Data Format A CITIfile package may contain more than one array of data. Arrays of data start after the BEGIN keyword, and the END keyword will follow the last data element in an array. A CITIfile package does not necessarily need to include data arrays; for instance, CITIfile could be used to store the current state of an instrument. In that case the keywords VAR, DATA, BEGIN, and END would not be required.
Understanding the CITIfile Data Format The CITIfile Data Format NAME DATA #NA REGISTER 1 VAR FREQ MAG 10 DATA S[1,1] RI SEG_LIST_BEGIN SEG 1000000000 4000000000 10 SEG_LIST_END BEGIN 0.86303E-1,-8.98651E-1 8.97491E-1,3.06915E-1 -4.96887E-1,7.87323E-1 -5.65338E-1,-7.05291E-1 8.94287E-1,-4.25537E-1 1.77551E-1,8.96606E-1 -9.35028E-1,-1.10504E-1 3.69079E-1,-9.13787E-1 7.80120E-1,5.37841E-1 -7.78350E-1,5.
Understanding the CITIfile Data Format The CITIfile Data Format #NA DUPLICATES 0 #NA ARB_SEG 1000000000 1000000000 1 #NA ARB_SEG 2000000000 3000000000 3 VAR_LIST_BEGIN 1000000000 2000000000 2500000000 3000000000 VAR_LIST_END BEGIN 1.12134E-3,1.73103E-3 4.23145E-3,-5.36775E-3 -0.56815E-3,5.32650E-3 -1.85942E-3,-4.07981E-3 END BEGIN 2.03895E-2,-0.82674E-2 -4.21371E-2,-0.24871E-2 0.21038E-2,-3.06778E-2 1.20315E-2,5.99861E-2 END BEGIN 4.45404E-1,4.31518E-1 8.34777E-1,-1.33056E-1 -7.09137E-1,5.58410E-1 4.
Understanding the CITIfile Data Format CITIfile Keywords CITIfile Keywords Keyword Explanation and Examples CITIFILE CITIFILE A.01.01 identifies the file as a CITIfile, and indicates the revision level of the file. The CITIfile keyword and revision code must precede any other keywords. The CITIfile keyword at the beginning of the package assures the device reading the file that the data that follows is in the CITIfile format. The revision number allows for future extensions of the CITIfile standard.
Understanding the CITIfile Data Format CITIfile Keywords SEG_LIST_BEGIN SEG_LIST_BEGIN indicates that a list of segments for the independent variable follow. Format for the segments is: [segment type] [start] [stop] [number of points]. The current implementation only supports a single segment. If there is more than one segment, the VAR_LIST_BEGIN construct is used. CITIfile revision A.01.00 supports only the SEG (linear segment) segment type.
Understanding the CITIfile Data Format CITIfile Keywords CONSTANT CONSTANT [name] [value] allows for the recording of values which don’t change when the independent variable changes. CONSTANTs are part of the main CITIfile definition. Users must not define their own CONSTANTs. Use the #KEYWORD device specification to create your own KEYWORD instead. The #NA device specification is an example of this. No constants were defined for revision A.01.00 of CITIfile. CITIfile revision A.01.
Understanding the CITIfile Data Format Useful Calculations Useful Calculations This section contains information on computing frequency points and expressing CITIfile data in other data formats. Computing Frequency Points In CITIfile, the frequency data is not listed point by point, only the start and stop values are given. If you are using a spreadsheet program, you can create a new frequency column to the left of the data pairs.
Understanding the CITIfile Data Format Useful Calculations Expressing CITIfile Data in Other Data Formats CITIfile data is represented in real and imaginary pairs. Equations can be used to express this information in logarithmic magnitude, phase, polar, and Smith chart formats. Refer to the following table for these equations.
Understanding the CITIfile Data Format Useful Calculations Example Data This example shows how the following CITIfile data for a three-point trace can be expressed in other data formats. CITIFILE A.01.00 #NA VERSION HP8753E.07.12 NAME DATA VAR FREQ MAG 3.0000 DATA S[11] RI SEG_LIST_BEGIN SEG 1550000000 1570000000 3.
Understanding the CITIfile Data Format Useful Calculations 9-14
10 Determining System Measurement Uncertainties 10-1
Determining System Measurement Uncertainties Introduction Introduction In any measurement, certain measurement errors associated with the system add uncertainty to the measured results. This uncertainty defines how accurately a device under test (DUT) can be measured. This chapter describes how the various network analyzer measurement error sources contribute to uncertainties in the magnitude and phase measurements of both transmission and reflection.
Determining System Measurement Uncertainties Sources of Measurement Errors Sources of Measurement Errors Measurement errors are made up of systematic errors, random errors, and drift errors. Each of these measurement error types is discussed in this section. Sources of Systematic Errors The residual (after measurement calibration) systematic errors result from imperfections in the calibration standards. All measurements are affected by dynamic accuracy.
Determining System Measurement Uncertainties Sources of Measurement Errors Sources of Random Errors The random error sources are noise, connector repeatability and interconnecting cable stability. There are two types of noise in any measurement system: low level noise (noise floor) and high level noise (trace noise). Low level noise is the broadband noise floor of the receiver which can be reduced through averaging or by changing the IF bandwidth.
Determining System Measurement Uncertainties Determining Expected System Performance Determining Expected System Performance Improper connection techniques and contact surfaces can degrade measurement accuracy. Proper connection techniques include using a torque wrench with proper torque limits, ensuring that the connector pin depths meet specifications, ensuring that the center conductor of sliding loads is properly set, and observing proper handling procedures for beadless airlines.
Determining System Measurement Uncertainties Determining Cable Stability Terms (CR1, CR2, CTM1, CTM2, CTP1, CTP2) Determining Cable Stability Terms (CR1, CR2, CTM1, CTM2, CTP1, CTP2) Cable stability is dependent on the cable used and the amount of cable movement between calibration and measurement.
Determining System Measurement Uncertainties Determining Cable Stability Terms (CR1, CR2, CTM1, CTM2, CTP1, CTP2) Figure 10-2 Cable Stability with a Short Connected Figure 10-2 and Figure 10-3 demonstrate the concepts useful in determining cable transmission stability. A short is connected to the free end. The DATA/MEM feature provides an indication of the two-way cable transmission stability.
Determining System Measurement Uncertainties Measurement Uncertainty Equations Measurement Uncertainty Equations Any measurement result is the vector sum of the actual test device response plus all error terms. The precise effect of each error term depends on its magnitude and phase relationship to the actual test device response. When the phase of an error response is not known, phase is assumed to be worst case (–l80° to +180°). Forward Reflection Uncertainty Equation 10-1.
Determining System Measurement Uncertainties Measurement Uncertainty Equations Forward Transmission Uncertainty Equation 10-3.
Determining System Measurement Uncertainties Measurement Uncertainty Equations Reverse Reflection Uncertainty Equation 10-5.
Determining System Measurement Uncertainties Measurement Uncertainty Equations Reverse Transmission Uncertainty Equation 10-7.
Determining System Measurement Uncertainties Measurement Uncertainty Equations 10-12
Index Numerics 6 GHz operation, Option 006, 7-3 75 Ohm impedance, Option 075, 7-4 A accessories available, 7-5 keyboard template, 7-12 measurement accessories, 7-5 adapters, 7-9 allocation, memory, 8-12 amplifier, 7-10 Analyzer panels front, 3-2 analyzer display, 3-5 analyzer functions, 5-4 analyzer options available, 7-3 6 GHz operation, Option 006, 7-3 75 Ohm impedance, Option 075, 7-4 configurable test set, Option 014, 7-3 harmonic mode, Option 002, 7-3 high stability frequency reference, Option 1D5, 7-
Index performance, determining, 10-5 power meters, 7-10, 7-11 preset state, 8-2 probe, 7-10 R rack mount flange kit with handles, Option 1CP, 7-4 rack mount flange kit without handles, Option 1CM, 7-4 random errors, 10-2, 10-4 raw error terms, 10-2 rear panel features and connectors, 3-10 receiver configuration, Option 011, 7-3 repeatability, connector, 10-5 repeatable errors, 10-2 residual errors, 10-2, 10-3 reverse reflection uncertainty, 10-10 transmission uncertainty, 10-11 RF ECal modules, 7-7 RF limi