Technical data

1-32
Installing the 86100C
To use the removable hard disk option
To exchange the removable hard disks
If you are removing the 86100C from a secure area, perform the procedure “To
declassify instrument memory” on page 1-30 before starting the following
steps.
1 Turn the 86100C on.
2 If the instrument has been returned from Agilent with a new mainframe
calibration, you must save a copy of the calibration file from the current hard
disk. Later in this procedure, you will install this file on the secure hard disk.
a On Calibrate menu, click All Calibrations. Click the Mainframe and
Skew tab. Record the calibration date for use later in this procedure. See
Figure 1-24.
Calibration date: _______________________________
Figure 1-24. Mainframe timebase calibration status and date
b Close the All Calibrations dialog box.
c On the File menu, click Exit to close the DCA application.
status message
calibration date