Technical data
1-24
Installing the 86100C
Step 11. Avoid costly repairs
Step 11. Avoid costly repairs
Figure 1-18. Avoid Costly Repairs
• Refer to the “Connector Care” on page 1-37, before making any electrical or op-
tical input connections.
CAUTION Electrical channel input circuits and the trigger input circuit can be damaged
by electrostatic discharge (ESD). Therefore, avoid applying static discharges
to the front-panel input connectors. Prior to connecting any coaxial cable to
the connectors, momentarily short the center and outer conductors of the
cable together. Avoid touching the front-panel input connectors without first
touching the frame of the instrument. Be sure that the instrument is properly
earth-grounded to prevent buildup of static charge. Wear a wrist-strap or heel-
strap. More information can be found in the online help.
CAUTION Optical channel fiber-optic connectors on plug-in modules are easily damaged
when connected to dirty or damaged cables and accessories. When you use
improper cleaning and handling techniques, you risk expensive instrument
repairs, damaged cables, and compromised measurements. Before you connect
any fiber-optic cable to the 86100C, refer to information in the online help.