Technical data
1-12
Installing the 86100C
Step 6. Connect the power
CAUTION This instrument has autoranging line voltage input. Be sure the supply voltage
is within the specified range.
CAUTION Windows XP registry: If the Infiniium DCA is mounted in a rack or cabinet, do
not use the system power switch to disconnect power from the instrument.
Instead, use the Infiniium DCA front-panel power switch. Using the system
power switch may corrupt the Windows XP registry requiring you to perform
the recovery procedure to restore normal operation to the Infiniium DCA.
Refer to “To recover the 86100C Drive-C” on page 2-25.