Firmware Release Notes
A.06.01 Release Notes, September 21, 2006 (Differences from Re-
lease A.06.00)
New Feature List
Added eye masks for 10G BASE LRM, 1000Base-LX10, 100Base-BX10, ITU g.984.2
General Defects Fixed
Corrected the 100Base-BX/LX mask scale to now have Delta T set to 8E-9 seconds
Function<N>:Vertical? Now returns correct value.
Differential TDR deskew is correctly locked out when calibration is active.
Missing specification for max input for 83496A included in help file.
Missing note on specification for 83491/2/3 included in help file.
Changed module vertical calibration messages for 86118A modules to say "Connect all
necessary cables and adaptors for measurement to the <Left | Right> module.
Addressed MEAS:TDR:MAX, MIN, AVE queries after default setup causing a crash.
No longer issues a GPIB hardware missing error using autoscale after tdr/tdt cal.
LTER? Now correctly sets bit 0 (COMP) when acquisition is complete.
:TDR:STIMulus:EXTernal:POLarity? now returns correct data when longform is on.
Fixed a problem when reading in a pattern waveform csv file written by hand.
Fixed intermittent application crash during certain TDR mode recall setups.
Fixed crash that happened with :MEAS:VTIM? .000001 in TDR mode right after default
setup.
Fixed a bug causing probe calibrations to fail (Traces may be displayed during
calibrations).
Added special options H93 and H95 for 5.00Gb/s and 6.25Gb/s to the 86105C.
Software Option Packages (#200 and #202) Defects Fixed
Fixed problems with jitter plots after default setup (#200).
Fixed problem where PJ is greater than DJ in some cases. (Still possible for PJ to be
greater than DJ if DJ/PJ small enough since they are calculated from different histograms)
( #200).
Added new S-parameter specs to TDR (#202).
Fixed defect preventing the save of differential or common mode S-Parameters from
remote (#202).
Known Issues
PLTS customers should not upgrade to A.06.00 and above software at this time since the
executable properly. PLTS customers should use DCA firmware version A.04.11.
86119A (Optical Sampling Oscilloscope) is designed to work on software versions A.03.06
and A.03.05 only.
86100A/B Mainframes are not supported on DCA Firmware Revision A.06.00 and above.