Technical data

11
86100A/B/C
Firmware Release Notes
86100C/86100B/86100A Firmware Release Notes
A.05.00 Release Notes, February 16, 2005 (Differences from Release
A.04.20)
New Feature List
Support for 86100A/B/C Mainframe
Maximum/Minimum/Average Measurements in TDR Mode
Export of waveform data in X & Y format
Display of ER Cal factor
Added the following masks (100BASE-BX_LX10.msk, SATA 1.5Gb TX 250 cycles.msk,
SATA 1.5GB TX 5 Cycles.msk)
General Firmware Defects Fixed in A.05.00
86100B + 54754A previously did not invert white trace color on print. The waveform color
is set to gray during printing to prevent a white waveform trace on a white background.
Repaired 86106B and other modules with un-labeled filter rates not successfully
completing Autoscale when filter is enabled
Phase is correct now for reference clock frequencies between 25 GHz and 39 GHz on the
40 GHz input of an 86107A Option 40 module.
Repaired inconsistent pattern waveform loading behavior
Extinction Ratio calibration factor is always loaded into dialog boxes
Removed stray entry out of range message at end of ER calibration
After 86100C power cycle, TDR mode is recalled properly
Removed the ability to load and store TDR/TDT calibration data since not all instrument
states were being saved properly.
Allowed TDT normalization to work with external TDR stimulus
Repaired a crash if remote commands (:MEAS:TDR: MAX MIN AVE) was sent without
putting the oscilloscope in TDR mode first.
Known Issues
PLTS customers should not upgrade to 5.0 software at this time since the executable will
fail to run properly. PLTS customers should use DCA firmware version 4.11.
86119A (Optical Sampling Oscilloscope) is designed to work on software versions A.03.06
and A.03.05 only.
A.04.20 Release Notes, June 9, 2005 (Differences from Release
A.04.11)
New Feature List
Support for 83496A Clock Recovery Module
Supports Picosecond Pulse Lab 4020 Single ended TDR sources and Picosecond Pulse Lab
4022 Differential TDR/T Sources. Provides external common and differential mode