Technical data

4-34
Replacing Instrument Assemblies
To Remove the A13 Acquisition (Option 001)
Figure 4-52. Removing the SW2 Cable from P19 Connector
6 Remove the board, as shown in Figure 4-53.
Figure 4-53. Removing the A13 Acquisition Board
7 Disconnect all of the remaining cables.
8 Refer to Figure 4-54 and Figure 4-55 for cable connections.
NOTE The left and right strobe cables are interchangeable.
P/O SW2