Technical data

4-9
Replacing Instrument Assemblies
To Remove the Mainframe Front Panel
Figure 4-9. Removing Screws that Secure the Cal Connector
6 Disconnect the W5 and W6 ribbon cables, shown in Figure 4-10.
Figure 4-10. Removing the W5 and W6 Ribbon Cables
7 Disconnect the W3 mylar flex cable, shown in Figure 4-11. Pry up the retainer slightly at either
end of the connector, using a small flat-blade screwdriver. Do not force the retainer; it should
remain attached to the body of the socket.
W5