Technical data

2-24
Performance Verification
Performance Test Record
Performance Test Record
Table 2-5. Performance Test Record
Agilent 86100A/B Series Mainframe
Firmware installed at time of calibration
(Software Revision):_________________
Serial Number _______________________
Recommended Test Interval – 1 Year/2000
hours
Recommended next testing _______________
Tested by ______________________
Work Order No. __________________
Date __________________________
Calibration temperature (from instrument)
_____________
Test Results
Time Interval Accuracy
Performance Test
Delay
Position
Settings (ns)
Min Actual Max
24 41.95 _______________
_______________
_______________
_______________
58.05
27.95 41.95 58.05
35.95 41.95 58.05
59.95 41.95 58.05
Period Accuracy
Scale Setting
Synthesized CW
Generator Setting
Min Actual Max
20 ps/div 10 GHz 91.9 ps ________ 108.1 ps
50 ps/div 5 GHz 191.8 ps ________ 208.2 ps
100 ps/div 2 GHz 491.5 ps ________ 508.5 ps
200 ps/div 1 GHz 0.9910 ps ________ 1.009 ns
500 ps/div 500 MHz 1.99 ns ________ 2.010 ns
1 ns/div 200 MHz 4.987 ns ________ 5.013 ns
2 ns/div 100 MHz 9.982 ns ________ 10.018 ns
5 ns/div 50 MHz 19.972 ns ________ 20.028 ns
10 ns/div 20 MHz 49.942 ns ________ 50.06 ns