Technical data
2-22
Performance Verification
Trigger Verification (Standard and Option 001)
Figure 2-16. Option 001 Trigger Sensitivity
2 On the front panel, press the Default Setup hardkey.
3 Turn on the electrical channel 2.
4 Turn all other channels off by pressing the corresponding hardkey.
5 On the DCA screen, press Trigger Level and set the Trigger Bandwidth to 2-12 GHz.
6 On the DCA screen, press Time Delay, set the reference to center screen, set the Scale (horizontal)
to 500 ps/div, and set the Delay From Trigger to 26.5 ns.
7 On the DCA screen, press Channel 2 Scale Offset and set the Scale (vertical) to 30 mV/div and
Offset to 0 V.
8 On the DCA screen, press Time then Jitter RMS.
9 On the DCA screen, press Amplitude, V p-p, and select Channel 2 as a source.
10 Set the Synthesized CW Generator FREQ and POWER LEVEL to the settings shown in Table
2-4, Column 1 and Column 2.
11 Make sure Channel 2 ON and that Channel 1 is OFF.
12 On the DCA screen, press Time Delay and temporarily set the Scale (horizontal) to approximately
100 ps/div in order to adjust the power level in the following step.
13 Adjust the Synthesized CW Generator’s POWER LEVEL so that the reading on the DCA Channel
2 is 200 mV p-p ± 1 mV.
14 Turn Channel 2 OFF and turn Channel 1 ON.
15 On the DCA, set the Scale (horizontal) as shown in Table 2-4, Column 3 for the FREQ in Column
1.
16 On the DCA screen, press Scale Offset and set the Channel 1 Scale (vertical) to 5 mV/div.
17 On the DCA screen, press Time Delay and set the Delay From Trigger to 26 ns.