Performance Verification
Jitter Performance Test
Figure 2-13. Rising Signal Intersecting with Center of Screen
20 On the DCA screen, press Measure, Histograms, Histogram Windowing.
21 Adjust the four marker knobs as follows:
Marker 1 Full Left Screen
Marker 2 +300
µV
Marker 3 Full Right Screen
Marker 4 –300
µV
22 Close the Histogram Windowing Dialog Box and set the Histogram State to ON.
23 The display should appear as shown in Figure 2-14.
Figure 2-14. Histogram Setup