Technical data
2-10
Performance Verification
Time Interval Accuracy Performance Test
Table 2-3. Period Accuracy for Scale Settings from 20 ns to 20 µs
Horizontal Scale Setting
Function/Arbitrary Waveform Generator
Settings
20 ns/div 10 MHz
50 ns/div 5 MHz
100 ns/div 2 MHz
200 ns/div 1 MHz
500 ns/div 500 kHz
1.0
µs/div
200 kHz
2.0
µs/div
100 kHz
5 µs/div
50 kHz
10 µs/div
20 kHz
20 µs/div
10 kHz