Technical data

2-7
Performance Verification
Time Interval Accuracy Performance Test
16 On the DCA screen, press Time Delay to set the Delay. See Figure 2-6. After changing the delay,
press the Clear Display hardkey before making the next measurement.
Figure 2-6. Delay Settings for Measuring Time Interval Accuracy
Period Accuracy for Scale Settings
17 On the DCA screen, press Time Delay and set Delay From Trigger to 24 ns.
18 On the DCA screen, press Setup, the channel in use, then press Advanced. See Figure 2-7.
Table 2-1. Period Accuracy for Delay Settings
Delay Setting in ns
24
27.95
35.95
59.95