Technical data

2-5
Performance Verification
Time Interval Accuracy Performance Test
Figure 2-2. Setting Time and Delay for Measuring Time Interval Accuracy
7 Ensure that the correct channel is enabled.
8 Turn all other channels off.
9 On the DCA screen, press Scale Offset for the channel in use, then set the scale (vertical) to
60 mV/div and the offset to 0 V. See Figure 2-3. Close dialog box.
Figure 2-3. Setting Scale and Offset for Measuring Time Interval Accuracy
10 On the DCA screen, press Setup, Acquisition, Enable Averaging, and set the number of averages
to 64. See Figure 2-4. Close dialog box.