Technical data

2-3
Performance Verification
Time Interval Accuracy Performance Test
Time Interval Accuracy Performance Test
The horizontal time interval accuracy of the mainframe and plug-in modules are compared to a
known reference.
Equipment Required
Period Accuracy for Delay Settings
Procedure
1 Warm up the system for at least 60 minutes, then perform a vertical calibration.
2 Connect the equipment as shown in Figure 2-1.
Equipment Critical Specifications Recommended Model/Part
Plug-in Module 54750A, 82480A, or 86100A series
plug-in with electrical input
Agilent 54751A, 83483A,
86112A
Synthesized CW
Generator
No substitution Agilent 83712B
Required Options:
1E5 High Stability Timebase
Recommended Options:
1E1— Output Step
Attenuator
1E8 — 1 Hz Frequency Res.
1E9 — 3.5 mm RF Out
Connector
Timing Generator No substitution Agilent 8133A standard or
Option 002
Function/Arbitrary
Waveform Generator
10 MHz square wave, 120 mV output,
stability 0.05 ppm/yr
Agilent 3325B Option 001
or Agilent 33250A
Transition Time Converter 2000 ps Agilent 15438A
Cable, BNC
50
, 122 cm (48.in)
Agilent 10503A
Adapter, BNC (m) to
SMA (m)
50
Agilent E9633A BNC (m) to
SMA (m), 50
Cable Assembly, 3.5 mm
(m) to 3.5 mm (m) (2
each)
61 cm (24 in); Frequency range of dc
to 26.5 GHz
Agilent 11500E
Adapter, 3.5 mm (f) to 3.5
mm (f) (2 each)
dc to 34 GHz Agilent 1250-1749