Specifications
22
Remote operation
The standard 8568B operates on the interface bus (GP-IB).
All analyzer control settings (with the exception of VIDEO
TRIGGER LEVEL, FOCUS, ALIGN, INTENSITY, FREQ ZERO, AMPTD
CAL, and LINE power) are remotely programmable. Function
values, marker frequency/amplitude, and traces may be output;
CRT labels and graphics may be input.
LCL
Returns analyzer to local control, if not locked out by controller.
Service request
SHIFT r calls an GP-IB request for service.
Interface codes
SH-l, AH1, T6, L4, SR1, RL1, PPO, DC1, DT1, CO, E1.
Options
All specifications are identical to the standard 8568B except as noted.
75 Ω input impedance (Option 001)
RF input #1
100 Hz to 1500 MHz, 75 Ω, BNC connector; dc coupled,
not fused.
Average noise level (0 dBm input attenuation, 1 Hz video filter)
Noise level displayed on RF input #1 –129 dBm with 10 Hz
resolution bandwidth, frequencies >1 MHz; < –106 dBm
for frequencies ≤ 1 MHz, but > 500 Hz.
Residual responses (no signal at input)
< –99 dBm, input #1.
400 Hz power line frequency operation (Option 400)
Line related sidebands
> 75 dB below peak of CW signal.
Residual responses (no signal at input)
< –95 dBm for frequencies > 500 Hz: < –105 dBm for
frequencies > 2.5 kHz. 0 dB input attenuation.
Power requirements
400 Hz ±10% line frequency; 100 or 120 volts (+5%, –10%)
line voltage; 50 to 60 Hz power line frequency for service only,
not for extended periods.
Temperature range (operating)
0 ° to 55 °C. Restricted to 0 ° to 35 °C, 50 Hz to 60 Hz.
Mounting kits
Rack flange kit (Option 908)
Rack flange kit to mount instruments with handles (Option 913)
Rack mount slide kit (Option 010)
Meets EIA RS310-C Specification.
Extra manuel (Option 910)
Retrofit kit (8568A +01 K)
Contains hardware and documentation to change an
85680A to an 85680B. Also includes BASIC operation
verification software for use with 9000 Series computers,
one 5 1/4-inch flexible disk, and one 3 1/2-inch micro
flexible disk.
Part numbers
Agilent 8568B spectrum analyzer
Option 001: 75 Ω (BNC), 100 Hz to 1500 MHz RF input #1
Option 400: 400 Hz power line frequency operation
Option 908: rack flange kit
Option 913: rack flange kit to mount instruments with handles
Option 010: rack mount slide kit (meets EIA RS310-C specification)
Option 910: extra manual
Accessories
Agilent 85650A quasi-peak adapter
Agilent 85685A RF preselector
Agilent 8568A +01 K retrofit kit
Agilent 8444A Option 059 tracking generator
Agilent 8447D preamplifier, 0.1 to 1300 MHz
Service kit for 8568B, P/N 08568-60005
Transit case
RF section P/N 1540-0602
IF section P/N 1540-0663
NOTE:
Dimensions in millimetres and (inches)
(Allow 100 mm, 4 inch clearance at
rear panel for interconnect cables.)
Specifications
See definition of specifications and supplemental characteristics on page 16.