Specifications

16
Frequency
Measurement range
100 Hz to 1500 MHz dc coupled and
100 kHz to 1500 MHz ac coupled.
Resolution
3 dB bandwidths of 10 Hz to 3 MHz in a 1, 3, 10 sequence.
Spectral purity
Noise sidebands > 80dB below peak of CW signal at frequency
offsets 30 x resolution bandwidth setting, for resolution
bandwidths 300 Hz.
Accuracy
Frequency reference accuracy (aging rate)
< 1 x 10
-9
/day (2 x 10
-7
/yr.). After 30-day warmup.
Center frequency
±(2% of frequency span + frequency reference accuracy x
tune frequency + 10 Hz) using error correction.
Frequency span
Spans > 1 MHz, ±(2% of indicated separation between
two points +0.5% of span); spans 1 MHz, ±(5% of indicated
separation +0.5% of span).
Marker
Normal: Center frequency accuracy + frequency span
accuracy between the marker and center frequency.
Freq Count: Frequency reference accuracy x displayed
frequency ± frequency counter resolution (spans 100 Hz).
Amplitude
Measurement range
–135 dBm to +30 dBm or equivalent in dBmV, dBµV; 40 nV to 7 V.
Dynamic range
Spurious responses
Second harmonic distortion and third order intermodulation
distortion > 70 dB below signal levels –30 dBm at the
input mixer.
Average noise level
< –135 dBm in 10 Hz resolution bandwidth.
Accuracy
Calibrator uncertainty
±0.2 dB.
Frequency response uncertainty
±1.0 dB, 100 Hz to 1500 MHz.
Sweep
Sweep time
20 msec full span to 1500 sec full span. Zero frequency span,
1 µsec full sweep to 1500 sec full sweep.
Inputs
RF inputs
100 Hz to 1500 MHz, 50 dc coupled (BNC fused); and
100 kHz to 1500 MHz, 50 ac coupled (type N).
Max input level
ac: +30 dBm (1 watt) continuous power: 100 watts, 10 µsec
pulse into > 50 dB attenuation.
dc; 0 volts dc coupled input and ±50 volts for ac coupled input.
Attenuator
70 dB range in 10 dB steps.
Outputs
Display
X, Y, and Z outputs for auxiliary CRT display.
Recorder
Horizontal sweep output (X), video output (Y), and
penlift/blanking output (Z) to drive an X-Y recorder.
Remote operation
All analyzer control settings (with the exception of video trigger
level, focus, align, intensity, frequency zero, amplitude cal, and line
power) may be programmed via the interface bus (GP-IB).
Specifications
Performance summary
Specifications describe the instrument's warranted performance over the temperature range 0° to 55 °C (except where noted.) Supplemental characteristics
are intended to provide information useful in applying the instrument by giving typical, but non-warranted, performance parameters. These are denoted as
"typical," "nominal," or "approximately." Where specifications are subject to minimization with the error correction routine, corrected limits are given unless
noted. (The error correction routine is a built-in routine using the CAL OUTPUT signal and pressing SHIFT W then SHIFT X. The message CORR'D appears
in the CRT display when the correction is being applied.)