Specifications

‘lhble 5-1. Tests Performed
Test Number and Name
1. Input Attenuator Switching Check
2. IF Gain Uncertainty
3. Scale Fidelity (Log)
4. Scale Fidelity (Linear)
5. Log Scale Switching
6. Resolution Bandwidths
7. Impulse and Resolution Bandwidths
(Option 462 Impulse
Bandwidth)t
8. Line-Related Sidebands
9. Residual FM
10. RF Gain Uncertainty
11. Sweep Time Accuracy
12. Average Noise Level
13. Residual Responses
14. Frequency Span Accuracy
15. Frequency Response
16. Second Harmonic Distortion
17. Third Order Intermodulation Distortion
18.
Cal Output Amplitude Accuracy
1
Equipment Required
HP 3335A
HP 3335A
HP 3335A
HP 3335A
HP 3335A
none
HP 8112A or HP 8116A
HP 3335A
none
none
none
none
HP
909A,
Option 012
HP
909A,
Option 012
HP 3335A or
HP
83640A*
HP 3335A and
HP
83640A*
or HP
8444A,
Option 059
HP 8902A or HP 436A
or HP 438A
HP 8482A
HP 11667A
HP 3335A
50 MHz LPF
(HP p/n 0955-0306)
HP 83640A’
HP 8335A
HP 8721A
50 MHz LPF
(HP p/n 0955-0306,
2 required)
HP 436A, Option 022
HP 8482A
Alternate sources are: HP
83620A,
HP 83623A, HP
83650A,
HP
8340A/B,
and
8341A/B.
i
Option 462 6
dB
Bandwidth instruments must be manually tested. See note.
Operation Verification 5-3