Specifications
Function
Descriptions
Sweep
Free Run
Line
Video
Characteristics
Sweep triggered by internal source.
Sweep triggered by power line frequency.
Sweep triggered by detected waveform of input signal at
an
adjustable level; signal must be
20.5
div peak-to-peak.
For sweeps of 10 ms and less (zero span) the signal must
have
External
>40
Hz rate.
[W)
(VIDEO_)
(KSy)
allows any envelope
rate,
but display will blank between triggers when sweep is
<20
ms.
Sweep triggered by rising edge of signal input to
rear-panel
BNC connector; trigger source must be
>2.4
V (10 V
maximum).
For sweep of 10 ms and less (zero span), trigger source
must
have
>40
Hz rate. (SHIFT]
IEXT_)
(KSx)
allows any trigger
source rate, but display will blank between low repetition
rate
trigger when sweep is
~20
ms.
Continuous Sequential sweeps initiated by the trigger.
Single
Single sweep armed on activation and initiated by trigger
(sweep
220
ms only).
Sweep Time
Range
Span = 0
1 ps to 10 ms in 1, 2, 5 sequence.
HZ
20 ms to 1500
s
in 1, 1.5, 2, 3, 5, 7.5, 10 sequence.
Span > 0
20 ms to 1500
s
in 1, 1.5, 2, 3, 5, 7.5, 10 sequence.
HZ
Characteristics 4-9