Specifications

Readout Accuracy
Span
2100
Hz
*(2% of frequency span + frequency reference error x tune
frequency + 10 Hz) in AUTO resolution bandwidth after adjusting
FREQ ZERO at stabilized temperature, and using the error-correction
function,
[SHIFT]
[FREQUENCY SPAN] (KSW). Add 30% of the resolution
bandwidth setting if error correction is not used.
Zero Frequency Span
Accuracy
Resolution Frequency Reference Error x
Readout
Bandwidth
Tune Frequency + Resolution
10 Hz to 300 Hz
10 Hz 1 Hz
1 kHz to 3 kHz
100 Hz 10 Hz
10 kHz to 3 MHz
1
kHz
100 Hz
Frequency Span
Range
0 Hz and 100 Hz to 1500 MHz over lo-division CRT horizontal axis;
variable in approximately 1% increments. Step keys change span in a
1, 2, 5 sequence.
Full Span
0 to 1500 MHz is immediately executed with the
(j-1
or
&NSTR PRESET] keys.
Frequency Span Readout Accuracy
Accuracy
Range
100 Hz to 1500 MHz
Readout Accuracy
Spans 21 MHz
f5%
of the actual frequency separation between
two points
+
0.5% span
Spans
>l
MHz
f2%
of the actual frequency separation between
two points + 0.5% span
Start or Stop
Center frequency readout accuracy +
l/2
Frequency Readout frequency span accuracy
Accuracy
3-2 Specifications