Specifications
3
Specifications
Introduction
Unless noted, all specifications are for AUTOCOUPLED FUNCTION
operation. Where specifications are subject to minimization with the
error-correction routine, corrected limits are given unless noted.
Nominal values provide useful, but nonwarranted, information about
functional performance.
Frequency
Measurement Range
100 Hz to 1500 MHz through two RF inputs: 100 Hz to 1500 MHz
dc coupled, and 100 kHz to 1500 MHz ac coupled.
Displayed Values
Frequency Reference Error and Accuracy
Accuracy
Aging rate
<I
x
10hg/day
and
~2.5
x
10F7/year
Temperature stability
~7
x 10eg over 0
“C
to 55
“C
range (25
“C
reference)
Note
Note
The term frequency reference error, when used later in this manual,
is defined as:
f
[aging rate/day x number of days since calibration
+
temperature stability].
When the analyzer is in STANDBY, the frequency reference
temperature is maintained at a steady state. For more information,
see “Warm-Up Time” under “General” in this chapter.
Note
Center Frequency
Changes in line voltage, gravitational field, and other environmental
conditions will affect the frequency reference accuracy.
Range
0 Hz to 1500 MHz
Specifications 3-1