Specifications

15. Check the fuse to make sure it is of the correct rating and type
for the ac line voltage selected. Fuse ratings for different line
voltages are indicated in the following table.
ac Line Voltage IF-Display Section RF Section
100/120
220/240
2
amperes FAST
BLO
2
amperes
FAST BLO
HP part number 2110-0002 HP part number 2110-0002
1
ampere
SLOW BLO
1
amperes
SLOW BLO
HP part number 2110-0007 HP part number 2110-0007
16. Insert the correct fuse, and push the fuse-pull lever into its
normal right-hand position.
17. Attach one slide inner-member bracket to each side of the
instrument using two M5x0.8~10 pan-head screws per side. See
callout
in 1 of Figure 2-13.
Figure 2-13. Attaching the Inner-Member Brackets
18. Mounting in an HP System Rack Cabinet:
a. Insert two Unistrut nuts,
callout
2 in Figure 2-13, into each of
the two vertical columns on the left side of the system cabinet.
Insert two Unistrut nuts into each of the two vertical columns
on the right side of the system cabinet.
b. Bolt an outer slide mount to the Unistrut nuts in each side of
the systems cabinet, using four
M5x0.8~12
pan-head screws per
side. See
callout
in 3 of Figure 2-13.
Installation