Specifications
16. Second
Harmonic
Distortion
Specification
Related Performance
Test
Description
For mixer level
s-30
dBm:
100 Hz to 10 MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <-60 dBc
>lO
MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <-70 dBc
Harmonic and Intermodulation Distortion
A 40 MHz signal is applied to the analyzer’s RF INPUT through a
50 MHz low-pass filter. The spectrum analyzer is adjusted to measure
the second harmonic at 80 MHz.
5.26
Operation Verification